Step 4    Select frequency of maximum

mains (line) rejection

Scroll though the main menu until

'Cond' is displayed and press 

P

which will result in the instrument
displaying '0000' with the most
significant digit flashing.   To prevent
accidental adjustment the Cond sub-
menu is protected by the code word
'SurE'' which should be entered

character by character using the 

Up

and 

Down

 buttons plus  

P

    to  move

control to the next digit.  Note: S is
entered as 5. When 'SurE' has been

entered pressing

  E

  will result in the

'FrE' prompt.  Press 

P

 to enter  the

function and using the 

Up

 or 

Down

buttons select '50'.  Finally, press 

E

twice to return to the 'Cond' prompt
in the main menu.

Step 5

Define function of P push-button

Select 'C--P' from the main menu

and press 

to reveal the function of

the

  P

 button in the operating mode.

Select percentage 'PC' and return to

the main menu by pressing 

 E

Step 6

Define rate display resolution

Select 'rESn' from the main menu

and press 

to reveal the current

resolution of the rate display.  Using

the 

Up

Down

 push buttons select a

resolution of '1' and return to the
'rESn' prompt in the main menu by

pressing 

 E.

Step 

7

Position rate & total decimal
points

Scroll though the main menu until

'd.P.' is displayed and then press

  P

.

The rate and total displays will be
activated with one digit of the total

display flashing.  Press the 

Up

 or

Down

 push-button until the second

least significant digit of the total
display and following decimal point
are flashing  i.e. a resolution of 0.1

Press 

P

 to move control to the rate

display. Using the 

Up

 or 

Down

 push-

button move the flashing digit to the
least significant position (right hand
side) which will result in no decimal
point being displayed.

Finally press 

E

 to return to the 'd.P.

prompt in the main menu.

Step 8

Calibrate the rate display

Scroll through the main menu until

'CAL' is displayed.  Press 

and the

instrument will request a 4mA input
by displaying  'ZErO'.  Set the input

current to 4.0000mA and press 

P

again which will reveal the existing

rate display.  Using the 

Up

Down

and  

P

 push-buttons enter the

required zero display of  0000

Press 

E

 to enter this calibration

figure.  The instrument will display
‘Ent’ for a few seconds before
returning to the  'ZErO' prompt.

Press the

 Up 

push-button

 

and the

instrument will request a 20mA input
by displaying  'SPAn'.  Set the input
current to 20.0000mA and again

press 

P

 which will reveal the existing

rate display at 20mA.  Using the 

Up

,

Down

 and 

P

 push-buttons enter the

required display of  1100  Press 

E

 to

enter this calibration figure. The
instrument will display ‘Ent’ for a few
seconds before returning to the
'SPAn' prompt.

Finally press 

E

 again to return to the

'CAL'  prompt in the main menu.

Step 9

Define the timebase

The rate display represents flow per
minute so it is necessary to set the
instrument timebase to 60.  This will
divide the rate display by 60 to
provide the correct total display.

Select 't-bASE' from the main menu

and press 

to reveal the current

setting.   Using the 

Up

 and 

Down

push buttons select 'tb - 60' and then
return to the main menu by  pressing

E.

Содержание BA554D

Страница 1: ...Issue 1 September 2000 BA554D 4 20mA loop powered field mounting rate totaliser...

Страница 2: ...centage clip off 5 15 Local reset of total display 5 16 Security code 5 17 Over and under range 6 Calibration Examples 6 1 Using an external current source 6 2 Using the internal references 7 Maintena...

Страница 3: ...ter Each time a 4 20mA current is applied to the instrument initialisation is performed After a short delay the following display sequence occurs All segments of the display are activated For about 3...

Страница 4: ...associated annunciators plus a flow indicator as shown on page 1 Rate Shows the flow rate on the smaller Display five digit display Total Shows the total flow since the display instrument was last re...

Страница 5: ...two mounting holes are located in a separate compartment with a sealed cover allowing the instrument to be installed without exposing the display assembly Terminals 2 and 4 are internally joined and...

Страница 6: ...stallation procedure 4 3 EMC The BA554D complies with the requirements of the European EMC Directive 89 336 EEC For specified immunity all wiring should be in screened twisted pairs with the screens e...

Страница 7: ...is displayed on the five digit display Using the instrument s CAL or SEt functions this display may be calibrated to show the flow rate in engineering units Both the zero display at 4mA and the span d...

Страница 8: ...cimal point clip off to adjust value Press or between 0 and 10 or and pressing Define Security Enter by Code to move to next digit Local reset or Fig 7B Programme Structure CLr Gtot E E E E E E E E P...

Страница 9: ...arly set the required full scale display accurate 20mA input current scale display using an Similarly set the required full to move to next digit pressing or and set required zero display by With accu...

Страница 10: ...ll new BA554D rate totalisers are supplied calibrated as requested at the time of ordering If calibration information is not supplied the BA554D will be set to display a rate of 0 00 with 4mA input an...

Страница 11: ...n and 4 20 to display current See section 5 12 Display Description of function SCALE t Total Scale Factor Defines the arithmetic relationship between the rate and total displays May be adjusted betwee...

Страница 12: ...ed press E to return to the main menu If the application requires flow rate to be the primary display i e shown on the large display the d P function enables the rate and total displays to be intercha...

Страница 13: ...ing the Up and Down buttons and the P button to move to the next digit enter the password SurE Pressing E will then give access to the sub menu Note S is entered as 5 5 9 1 AC rejection FrE WARNING If...

Страница 14: ...play rate display divided by Per second tb 1 1 Per minute tb 60 60 Per hour tb 3600 3600 Select the required divisor and press E to return to the main menu 5 12 Function of the P push button C P This...

Страница 15: ...flow may be adjusted in one percent increments using the Up or Down buttons When set as required press the E button to enter the revised figure and return to the main programme menu 5 15 Local reset o...

Страница 16: ...n using a traceable 4 20mA calibrator Section 6 2 contains an example of calibration using the instruments internal 4 and 20mA references which although not traceable are adequate for many industrial...

Страница 17: ...owing decimal point are flashing i e a resolution of 0 1 Press P to move control to the rate display Using the Up or Down push button move the flashing digit to the least significant position right ha...

Страница 18: ...ep 12 Reset the grand total to zero To reset the grand total to zero select CLr Gtot from the main menu and press P which will cause the instrument to display Clr no Toggle this to CLr YES using the U...

Страница 19: ...respond by displaying root which is the first function in the main menu Steps Perform as described in previous 3 to 7 calibration example in section 6 1 Step 8 Calibrate the rate display Scroll throug...

Страница 20: ...ble to enter the programme mode Incorrect security code entered Enter correct security code or fit security link in override position See Fig 8 7 2 Fault finding after commissioning ENSURE PLANT SAFET...

Страница 21: ...still does not function it is likely that the fault is within the protection components on the terminal assembly Terminal assemblies may be exchanged on site providing the replacement includes termina...

Страница 22: ...ed as high or low rate or total alarms with normally open or normally closed outputs Fig 9 illustrates the conditions available for a rate alarm and shows which are fail safe i e output is in the alar...

Страница 23: ...the alarm function without changing the alarm parameters See section 8 2 3 tYPE Type of alarm Defines whether the alarm operates on the rate or total display See section 8 2 4 SPr1 Alarm setpoint 1 o...

Страница 24: ...tYPE This function conditions the alarm to operate on the rate or total display Alarm 1 and Alarm 2 may both be conditioned as rate or total alarms or one may be conditioned for rate and the other fo...

Страница 25: ...buttons and the P button to move to the next digit When the required hysteresis has been entered press E to return to the alarm menu e g A BA554D calibrated to display a flow of 0 to 5000 with a high...

Страница 26: ...changed by operating the Up and Down push buttons and the P button to move to the next digit When the required code has been entered press E twice to return to the AcSP prompt in the Programme Menu Co...

Страница 27: ...isplay backlight Reducing the supply voltage below 18V or fitting a curent limiting resistor in series with the supply will reduce the display brilliance 8 4 External push buttons For applications req...

Страница 28: ...ault finding during commissioning 7 1 after commissioning 7 2 Flow indicator 2 2 Flow transmitter loops 3 1 Grand total 2 1 clear 5 3 5 10 Guarantee 7 5 HI LO 8 2 2 8 2 6 HStr 8 2 2 8 2 8 Internal ref...

Страница 29: ...struction manual Namely The rate display zero can not be elevated i e 4mA input current must correspond to zero flow 1 and 1 do not flash when the rate display overranges The timebase is identified in...

Отзывы: