9.4.3

Instrument function:   

FunCtion

The BA377E-SS may be configured as a Timer or as
a   Clock.     This   section   of   the   instruction   manual
describes   the   Clock   configuration,   for   details   of
Timer configuration see section 6.5.

To   reveal   the   existing   function   of   the   instrument
select 

FunCtion

  from   the   configuration   menu   and

press 

(

.     If 

CLoC

  is displayed, the instrument is

already configured as a Clock therefore press 

)

 to

return to the 

FunCtion

  prompt in the configuration

menu.   If 

ELAP5E

  is displayed, press the 

*

  or 

&

button to change the setting to 

CLoC

 followed by the

(

  button which will result in a   

0000

  prompt being

displayed   with   the   first   digit   flashing.     This   is   a
request   for   the   instruction   to   be   confirmed   by
entering 

5urE

  using the 

*

  or 

&

  button to adjust

the flashing digit and the 

(

 button to move control

to   the   next   digit.    When  

5urE

  has   been   entered,

pressing  

)

  will change the instrument to a Clock

and return the instrument to the display mode.   To
configure the Clock enter the configuration menu by
pressing the 

(

 and

 )

 buttons simultaneously until

function

 is displayed.  

 

9.4.4

Display format:   

di5PLAY

The BA377E-SS Clock may be configured to display
time   in   a   12   or   24   hour   format   with   or   without
seconds.   When a 12 hour format is selected, AM is
denoted by an  

A

  prefix at the left hand side of the

display, similarly a 

P

 is displayed to denote PM.

To check or change the display format of the clock,
select  

di5PLAY

  from   the   configuration   menu   and

press 

(

 which will reveal the existing setting which

can be changed by pressing the  

*

  or  

&

  button

followed by the 

)

 button to enter the selection and

return   to   the   configuration   menu.       The   options
available are shown below:

12

 

:

 

00

 

:

 

00

Twelve hour format with seconds

12

 

:

 

00

Twelve hour format without seconds

24

 

:

 

00

 

:

 

00

Twenty four hour format with seconds

24

 

:

 

00

Twenty four hour format without seconds

9.4.5

Set clock display time:   

5Et

This function sets the time displayed by the Clock
which   must   be  entered   each   time   the  BA377E-SS
Clock is powered.     Until a set time is entered the
Clock will display a flashing 

00

 

:

 

00

 

:

 

00

 or 

00

 

:

 

00

 for 24

hour formats, or  

A12 : 00 : 00

  or  

A12 : 00

  for 12 hour

formats.

To   adjust   the   clock   set   time,   select  

5Et

  from   the

configuration   menu   and   press  

(

  that   will   show

00

 

:

 

00

 

:

 

00

  or   the   existing   set   time   with   the   hours

flashing.     When   setting   the   clock   display   time
seconds are always shown.

Using the 

*

 or 

&

 button adjust the flashing hours

and then press 

(

 to transfer control to the minutes

display, pressing 

(

 again will transfer control to the

seconds display.     When the clock display time has
been adjusted press 

)

 to start the Clock and return

to the 

5Et

 prompt in the configuration menu.

9.4.6

Enter synchronising time:   

5YnC  t

When the external reset terminals RS1 and RS2 are
connected   together   the   clock   is   stopped   and   the
displayed   time   is   set   to   the   synchronising   time.
When the connection between the reset terminals is
removed, the clock restarts from the synchronising
time.  This function enables the BA377E-SS clock to
be synchronised with a master clock.

To enter the synchronising time, select  

5YnC t

  from

the   configuration   menu   and   press  

(

  which   will

reveal the existing synchronising time with the hours
flashing.       Using   the  

*

  or  

&

  button   adjust   the

hours and then press  

(

  to transfer control to the

minutes   display,   pressing  

(

  again   will   transfer

control   to   the   seconds   display.       When   the
synchronising time has been set as required, press

)

 to return to the configuration menu.

9.4.7

Control output 1 (Optional):   

oP1

Control   output   1   is   an   optional   factory   fitted
galvanically isolated solid state switch contact output
which can be configured to turn  

on

  (output closed)

and 

off 

(output open) twice in each twenty-four hour

period.  The control output status is shown by the 1
control output display annunciator.

Function 

oP1

  contains   a   sub-menu   with   five   sub-

functions:

EnbL

Number of times output 1
turns 

on 

&

 off 

in 24 hours.

oP 1 : on 1

First time oP 1 turns 

on

 (closes)

oP 1 : oFF 1

First time oP 1 turns 

off

 (opens)

oP 1 : on 2

Second time oP 1 turns 

on

 (closes)

oP 1 : oFF 2

Second time  oP 1 turns 

off

 (opens)

To check or change the function of Control output 1
select 

oP1

 from the configuration menu and press 

(

which will reveal the first sub-function 

EnbL.

Содержание BA377E-SS

Страница 1: ...Issue 6 4th May 2021 BA377E SS Rugged one input intrinsically safe Timer or Clock Issue 6...

Страница 2: ...ear sealing kit 6 OPERATION AS A TIMER 6 1 Initialisation 6 2 Controls when configured as a Timer 6 3 Displays when configured as a Timer 6 4 Timer structure 6 5 Configuration as a Timer 6 5 1 Accessi...

Страница 3: ...YnC t 9 4 7 Control output 1 Optional oP1 9 4 8 Enable Control output 1 EnbL 9 4 9 On and off times oP 1 on 1 oP 1 off 1 oP 1 on 2 oP 1 off 2 9 4 10 Control output 2 Optional oP2 9 4 11 Enable control...

Страница 4: ...plied with each instrument 2 INTRINSIC SAFETY CERTIFICATION The BA377E SS has IECEx and ATEX gas certification This section of the instruction manual describes ATEX gas certification IECEx and other a...

Страница 5: ...28V dc Ii 200mA dc Pi 0 84W Any certified Zener barrier or galvanic isolator with output safety parameters equal to or less than these limits may be used to power the BA377E SS The maximum equivalent...

Страница 6: ...unlikely to make a significant difference to the allowable cable parameters 2 5 2 Sensors that require energising Switch contacts proximity detectors and open collector sensors require energising whic...

Страница 7: ...application Almost any intrinsically safe relay with certification permitting the contacts to be connected to equipment in the hazardous area may be used When used as a Timer the BA377E SS may also b...

Страница 8: ...internally connected together as shown in Fig 15 If any of these terminals are earthed as shown in Figs 1 2 the other common terminals should only be connected to the same earth i e the barrier busbar...

Страница 9: ...e input must be maintained not momentary The BA377E SS input is not used when the instrument is configured as a Clock No Zener barrier is required in series with the input if the intrinsically safe se...

Страница 10: ...n Uo 28V dc Io 200mA dc Po 0 84W The BA377E SS Timer may therefore be directly connected to and controlled by most certified intrinsically safe high level voltage signals The maximum permitted cable p...

Страница 11: ...requirements for simple apparatus This allows the control output terminals A1 A2 and A3 A4 to be connected to almost any intrinsically safe circuit protected by a Zener barrier having output paramete...

Страница 12: ...the hazardous area although it may be difficult to find isolators for some types of sensor The BA377E SS input energising link should be fitted to suite the output of the galvanic isolator not the se...

Страница 13: ...hese requirements The BA377E SS contains a configurable debounce circuit to prevent false triggering of the instrument Three levels of debounce protection are available See section 6 5 6 3 2 4 Open co...

Страница 14: ...may be used but a galvanic isolator or IS relay is required to transfer the contact closure into the hazardous area Fig 5 illustrates how a BA377E SS Timer may be reset from both the safe and the haza...

Страница 15: ...ections 3 1 and 3 2 When correctly installed the BA377E SS Timer or Clock does not invalidate the certification of the Ex e Ex n or Ex p panel enclosure allowing installation with higher power control...

Страница 16: ...idate the Ex p panel s impact and ingress protection as the front of the Timer or Clock complies with Ex p impact and ingress requirements Although mounted in an Ex p panel enclosure the BA377E SS rem...

Страница 17: ...r Clock s captive gasket and ensure that it is not damaged before inserting the Timer or Clock into the panel enclosure aperture c If the enclosure panel is less than 1 0mm thick or is non metallic an...

Страница 18: ...wing the requested units of measurement if this information is not supplied when the instrument is ordered a blank card will be fitted A pack of self adhesive scale cards printed with common units of...

Страница 19: ...g the units of measurement specified when the instrument was ordered if the units are not specified a blank scale card will be fitted A pack of scale cards pre printed with common units of measurement...

Страница 20: ...er cycle can be accomplished via the front panel push buttons or by a remote contact A grand total time is maintained which can be viewed by operating the front panel push buttons It may be reset to z...

Страница 21: ...de See 6 5 15 Shows in succession firmware version number instrument function elap5e and any output accessories that are fitted A Dual Control Outputs Access to configuration menu 6 3 Displays when co...

Страница 22: ...delay time automatically returns to Running Complete Entered automatically from Running when there are no more timing cycles to perform When the CYCLE5 function is enabled the timing cycle can be rep...

Страница 23: ...e instrument using the or push button to adjust the flashing digit and the push button to transfer control to the next digit If the correct code has been entered pressing will cause the first paramete...

Страница 24: ...e input by linking terminals 3 4 See section 6 5 5 dEbounCE Defines the level of debounce applied to the input to prevent false functioning dEFAuLt HEAVY LiGHt See section 6 5 6 Display Summary of fun...

Страница 25: ...m the display mode to 5Et t and r5t dela See section 6 5 15 Display Summary of function uP or dn Direction of count Defines whether the Timer times down from 5Et t to zero or times up from zero to the...

Страница 26: ...ontrol output 1 See section 6 5 26 CLr Gtot Resets grand total to zero This function resets the grand total which represents the total Timer run time from within the configuration menu when CLr YE5 is...

Страница 27: ...as required press twice to return to the configuration menu or repeatedly press the or button until the required type of input is displayed and then press twice to return to the configuration menu One...

Страница 28: ...of count selected in the uP or dn function oFF Lower display disabled If the CYCLE5 function is enabled the following three options are available 5td Lower display shows the number of cycles requeste...

Страница 29: ...ess which will reveal the existing setting The required units can be selected by pressing the or button followed by the button to enter the selection and return to the configuration menu The options a...

Страница 30: ...30...

Страница 31: ...31...

Страница 32: ...o the next digit When set as required enter the selection and return to the cycl cnt prompt in the sub menu by operating the button 6 5 14 Restart delay r5t dela This sub function defines the time del...

Страница 33: ...tely Continue When power is restored the Timer will continue from where it stopped without any manual intervention To check or change the function select P fail from the configuration menu and press w...

Страница 34: ...figuration example in section 7 2 of this manual illustrate the use of the control outputs Further examples can be found in Application Guide AG374 which is available on the BEKA website 6 5 22 Contro...

Страница 35: ...n be independently configured to turn on and off when the Timer is any of it s five states It s functions and configuration are identical to control output 1 described in sections 6 5 21 to 6 5 25 The...

Страница 36: ...azardous area contact is closed The display is required in hours and minutes within the hazardous area and is to be reset to zero by a push button located in the same hazardous area not by the instrum...

Страница 37: ...n is operated Fig 19 Control of valve in a hazardous area The required instrument configuration for this example are shown below Function Display Setting Access code CodE 0000 Function FunCtion ELAP5E...

Страница 38: ...ct BEKA if code is lost 8 2 Fault finding after commissioning ENSURE PLANT SAFETY BEFORE STARTING MAINTENANCE Live maintenance is permitted on intrinsically safe equipment installed in a hazardous are...

Страница 39: ...e regularly checked Inspection frequency should be chosen to suit the environmental conditions 8 5 Guarantee Instruments which fail within the guarantee period should be returned to BEKA associates or...

Страница 40: ...tialisation the instrument will display a flashing display which is a request for the local time to be entered The clock will not start to function until a display time has been entered or the remote...

Страница 41: ...hroughout this manual push buttons are shown as and and legends displayed by the Clock are shown in a seven segment font as they appear on the instrument e g di5PLAY and CodE Access to the configurati...

Страница 42: ...42...

Страница 43: ...Optional Control output 1 can turn on output closed and off output open twice in each twenty four hour period This function contains five sub function EnbL Enables output 1 and defines if it turns on...

Страница 44: ...et time is entered the Clock will display a flashing 00 00 00 or 00 00 for 24 hour formats or A12 00 00 or A12 00 for 12 hour formats To adjust the clock set time select 5Et from the configuration men...

Страница 45: ...tered return to the configuration menu by pressing the button twice 9 4 10 Control output 2 Optional oP2 Control output 2 is an optional factory fitted output which can be configured to turn on output...

Страница 46: ...e display mode may be protected by a four digit alphanumeric security code which must be entered to gain access Default security code 0000 allows unrestricted access With ACCd displayed press to enter...

Страница 47: ...ng will then change the instrument to a Clock and return the display to FunCtion in the configuration menu Step 3 Select display format Using the or button select di5PLAY in the configuration menu and...

Страница 48: ...ss to the configuration menu Using the and buttons select CodE from the configuration menu and press which will reveal 0000 with the first digit flashing This example requires the security code to be...

Страница 49: ...377E SS fails after it has been functioning correctly as a Clock the following table may help to identify the cause of the failure Symptom Cause Check No display No power supply That there is between...

Страница 50: ...combustible dust in air is likely to occur occasionally in normal operation Zone 22 explosive atmosphere in the form of a cloud of combustible dust in air is not likely to occur in normal operation b...

Страница 51: ...ssible to mount the intrinsically safe interface in the same Ex t panel enclosure as the Timer or Clock if in normal operation the outside temperature of the enclosure is not affected The Timer or Clo...

Страница 52: ...the main section of this manual and both refer to the same standards Therefore the ATEX installation requirements specified in sections 2 3 and Appendix 1 of this manual may be used for IECEx installa...

Страница 53: ...y therefore also be used for US and Canadian installations subject to compliance with the local codes of practice ETL and cETL intrinsic safety codes CL I Div 1 Groups A B C D T5 US IS gas Div cert CL...

Страница 54: ...54...

Страница 55: ...55...

Страница 56: ...56...

Страница 57: ...57...

Страница 58: ...58...

Страница 59: ...59...

Страница 60: ...60...

Страница 61: ...61...

Страница 62: ...62...

Страница 63: ...63...

Страница 64: ...64...

Страница 65: ...65...

Отзывы: