Display
Summary of function
FiLtEr
Display filter
Is an adjustable digital filter to
reduce the noise on the rate display.
The filter has two parameters each
represented by a digit adjustable
between 0 and 9. The first digit
defines the amount of filtering
applied to the display, the second
digit the deviation from the displayed
rate at which the filter will be
overridden and the rate display will
move rapidly to the new value.
See section 6.20
&
or
*
Direction of count
Determines whether pulses at inputs
A and b increment or decrement the
total display.
See section 6.21
CLr VAL
Reset value
Defines a preset number to which
the total display will be set when the
BA368E Counter is locally or
remotely reset. Enables the
instrument to count down from a
preset number.
See section 6.22
LoC CLr
Local clear
Contains sub-menu with two
functions enabling the total and the
grand total to be reset via the front
panel push buttons while the
Counter is in the display mode.
See section 6.23
Clr tot
When
on
is selected total display is
reset when
&
and
*
buttons are
operated simultaneously for more
than 3 seconds in the display mode.
See section 6.24
Clr Gtot
When
on
is selected the grand total
may be reset when
)
and
*
buttons are operated simultaneously
for more than 10 seconds in the
display mode - see section 2.2 for
details.
Note:
Once reset, the grand total
can not be restored.
See section 6.25
Display
Summary of function
Clr Gtot
Resets grand total to zero from
within configuration menu.
This function resets the grand total
to zero from within the configuration
menu when
CLr YE5
is selected.
Note
: Once reset, the grand total
can not be recovered.
See section 6.26
CodE
Access code
Defines a four digit alphanumeric
code that must be entered to gain
access to the configuration menu.
Default code 0000 disables the
security function and allows
unrestricted access to all
configuration functions.
See section 6.27
r5Et dEF
Reset to factory defaults
Returns the BA368E Counter to the
factory defaults shown in section 6.0
To prevent accidental use the
request must be confirmed by
entering
5urE
before the reset will
be executed.
See section 6.28
19
Содержание BA368E
Страница 1: ...Issue 5 9th July 2019 BA368E Two Input Intrinsically safe Counter Issue 5 ...
Страница 20: ...20 ...
Страница 21: ...21 ...
Страница 46: ...46 ...
Страница 47: ...47 ...
Страница 48: ...48 ...
Страница 49: ...49 ...
Страница 50: ...50 ...
Страница 51: ...51 ...
Страница 52: ...52 ...
Страница 53: ...53 ...
Страница 54: ...54 ...
Страница 55: ...55 ...
Страница 56: ...56 ...
Страница 57: ...57 ...