![BEKA BA334NG Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/beka/ba334ng/ba334ng_manual_2723858023.webp)
23
6.18 Local reset:
LoC CLr
The Local reset function contains two sub-functions
Clr tot
and
Clr gtot
which when enabled allow the
total display and grand total to be reset to zero via
the instrument push buttons while the Rate Totaliser
is in the totalisation mode.
6.19 Local total reset:
Clr tot
Clr tot
is a sub-menu in the
LoC clr
function which
when activated allows an operator to reset the total
display to zero while in the totalisation mode by
operating the
&
and
*
push buttons
simultaneously for more than three seconds.
Select
loc clr
in the configuration menu and press
(
which will reveal the
clr tot
prompt then
operate
(
again which will show
if the local total
reset is
on
or
oFF
. If set as required operate the
)
button twice to return to the configuration menu, or
the
&
or
*
button to change the setting followed
by the
)
button twice to enter the change and
return to the
loc clr
prompt in the configuration
menu.
Note:
The total display may also be reset to zero remotely
by connecting terminals RS1 and RS2 together for
more than one second. See sections 3.7 and 4.3 of
this manual.
6.20 Local grand total reset:
clr gtot
The grand total is a separate sixteen digit counter
which is incremented in parallel with the total
display, but is not zeroed when the total display is
reset to zero. The grand total may be viewed in the
totalisation mode in two eight digit sections as
described in section 2.2 of this manual.
Clr gtot
is a sub-menu in the
loc clr
function
which when activated allows the operator to reset
the grand total display to zero in the totalisation
mode by operating the
)
and
*
push buttons
simultaneously for more than ten seconds.
Select
loc clr
in the configuration menu and press
(
which will reveal
clr tot
. Using the
&
or
*
button to select
clr gtot
and press
(
which will
show
if local grand total reset is
on
or
oFF
. If set as
required operate the
)
button twice to return to the
configuration menu, or the
&
or
*
button to
change the setting followed by the
)
button twice
to enter the change and return to the
LoC .clr
prompt in the configuration menu.
6.21 Grand total reset from configuration
menu:
Clr Gtot
The grand total is a separate sixteen digit counter
which is incremented in parallel with the total
display, but is not zeroed when the total display is
reset to zero. The grand total may be viewed in
the totalisation mode in two eight digit sections as
described in section 2.2 of this manual.
The grand total can be reset to zero from within the
configuration menu using this
CLr. Gtot
function, or
from the totalisation mode if sub-function
clr gtot
in the
loc clr
function is activated - see 6.20.
To zero the grand total from within the
configuration menu select
CLr Gtot
and press
(
which will cause the instrument to display
Clr . no
with
no
flashing. Press the
&
or
*
push
button until
Clr . YE5
is displayed and then press
(
which will result in a
0000
prompt being
displayed with the first digit flashing. This is a
request for the instruction to be confirmed by
entering
5urE
using the
&
or
*
buttons and the
(
button to move control to the next digit.
Pressing
)
will then reset the grand total to zero
and return the Rate Totaliser to the configuration
menu.
Note:
Once reset, the grand total can not be
recovered.
6.22 Security code:
CodE
Access to the instrument configuration menu may
be protected by a four digit security code which
must be entered to gain access. New instruments
are configured with the default security code
0000
which allows unrestricted access to all
configuration functions.
To enter a new security code select
CodE
from the
configuration menu and press
(
which will cause
the Rate Totaliser to display
0000
with one digit
flashing. The flashing digit may be adjusted using
the
&
or
*
push buttons, when set as required
operating the
(
button will transfer control to the
next digit. When all the digits have been adjusted
press
)
to return to the
CodE
prompt. The revised
security code will be activated when the Rate
Totaliser is returned to the totalisation mode.
Please contact BEKA associates sales department
if the security code is lost.