Function of the Buttons
1) 
Power on and off button.
To turn the TESTER on or off, keep the button pressed for 5 seconds.
2) Group of navigation buttons of the TESTER.
a) "DOWN" 
button by pressing this button the pages are scrolled downwards. (indication 6)
b) "ENTER" button Pressing the button on page 1 you access:
1) Model verification of the interrogated device
2) Matriculation number
3) Week and year of construction
4) Estimated residual working days
5) Ignition cycles
6) Detected alarms
7) Corrections made
The "ENTER" button on page 3 "PRINT", serves to confirm the start of printing
The "ENTER" button, on page 4 "TAR 175", is used to modify the TAR according to the needs, pressing it will note
the blinking of the writing, and pressing "UP" or "DOWN" increases or decreases the data.
The "ENTER" button on page 5 "NEW" is used to reset the device to the factory data by resetting the counters
The "ENTER" button on page 6 "SAVE" is used to save permanently edited data
c) "UP" button by pressing this button the pages are scrolled upwards. (see indication 6)
Checking the status of the battery.
Pressing the "UP" and "DOWN" buttons at the same time, the message BAT appears on the display followed by three
segments; Segments full, charged battery, empty segments empty battery.

DISPLAY
The display used on the TESTER is of the alphanumeric type and is used to display the following information:
3) "LEL" unit of measurement for explosive gas; "ppm" for the toxic gas, the exchange takes place automatically
according to the type of probe.
4) Percentage reading, indicates the quantity of gas lost in the environment.
N.B. The word "FAULT" replaces the reading of the percentage of gas in case of probe failure.
5) Sign of "%" lights up only when explosive gas is read.
6) Reading page number; The tester contains 1 to 7 pages, depending on the model.
7) "ASTERISK with a line" icon lights up when the tester is not connected to the probe in question.
8) Clock icon, indicates that the probe in question is in the "WARMUP" wait state.
9) "IR" transmission element given to the tamper.
10) "USB" port TS1008 connection with equipment to be examined

Edit

Enter

TS1008

Sensor

Analyzer

LEL  20

*

%

2

UP                                   Down

1

3

5

6 5

9

  Print  IR

1

2

7

Edit

Enter

TS1008

Sensor

Analyzer

LEL  20

*

10

3

4

5

6

%

2

UP                                   Down

8

Page 4

Components and commands

Содержание TS1008-V1

Страница 1: ... of the cycle Check and view how many times it has been turned on Check and view how many times it has detectedan Alarm Check and view how many corrections have been made Check and display the operating range of LEL Check and display the operation of the device Issues the receipt proving the operation TS1008 V1 Self certifying Tester for GAS Detectors and Sensors Certify the test by printing the r...

Страница 2: ...s new device is suitable to test all conventional probes to detect explosive and toxic gas manufactured by BEINAT Srl This device allows to read all of the data and the configuration of work that are in the memory of the probes also having supplied by the serial transmission it prints the ticket that confirms the testing data certifying your own work By the Tester TS1008 you can read all of the ev...

Страница 3: ...the 4 screws located at the back of the instrument 3 Raise the cohesion 4 Remove the batteries 5 Insert the new batteries positioning them in the correct direction 6 Replace the cover and tighten the 4 screws 7 Turn on the instrument and continue in your work Use Batteries must never be exposed to temperatures above 40 C To ensure optimal durability use them at room temperature If used at both low...

Страница 4: ...on 6 Checking the status of the battery Pressing the UP and DOWN buttons at the same time the message BAT appears on the display followed by three segments Segments full charged battery empty segments empty battery DISPLAY The display used on the TESTER is of the alphanumeric type and is used to display the following information 3 LEL unit of measurement for explosive gas ppm for the toxic gas the...

Страница 5: ...ctioning and is reserved for authorized technicians To see the next page press DOWN PAGE 2 to scroll the page press ENTER Displays the percentage of gas dispersed in the environment can be expressed in of LEL or in ppm In case of failure appears only FAULT Displays the type of gas for which the probe has been calibrated To see the next page press DOWN PAGE 3 PRINT to print point the IR transmitter...

Страница 6: ...rged if it is not recharge it If the device does not detect Check that the cables are well plugged in and make contact If the device does not print Check that the instrument is in line with the printer Check that the printer is turned on and is in line with the instrument In the event other problems arise directly contact a specialist and or authorised technician or your BEINAT S r l dealer Proble...

Страница 7: ...inder within the percentages described for each type of equipment Ordinary and extraordinary maintenance and putting out of service at the end of the operating period guaranteed by the manufacturer must be carried out by authorized and or specialized personnel LEL 20 Senso r Analyzer TS1008 Edit Enter UP Down Accessories To perform the test and to obtain a self certification it is necessary to use...

Страница 8: ...nd from its enforced suspension of use DISPOSAL OF OLD ELECTRICAL ELECTRONIC EQUIPMENT This symbol on the product or its packaging to indicates that this product shall not be treated as household waste Instead it shall be handed over to the applicable collection point for the recycling of electrical and electronic equipment such as for example sales points in case you buy a new and similar product...

Отзывы: