background image

 
 
 
 
 
 
 
 
 

 

Instructions for Fluorescent 

Immunoassay Analyzer Operation 

 
 
 
 
 
 

 

Beijing Dianotech Sci-Tech Co., Ltd 

Содержание FIC-Q100N

Страница 1: ...Instructions for Fluorescent Immunoassay Analyzer Operation Beijing Dianotech Sci Tech Co Ltd ...

Страница 2: ...RS OF ANALYZER 4 3 2 TRANSPORTATION AND STORAGE CONDITIONS OF ANALYZER 4 3 3 OPERATING CONDITIONS OF ANALYZER 4 CHAPTER IV ANALYZER INSTALLATION 5 CHAPTER V DETAILED OPERATING STEPS OF ANALYZER 5 5 1 SYSTEM SETUP 6 5 2 PROJECT MANAGEMENT 13 5 3 SINGLE SAMPLE TESTING 14 5 4 BATCH TESTING 17 5 5 HISTORIC RECORDS 20 CHAPTERVI MAINTENANCE ANDSERVICE 21 6 1 MAINTENANCE AND SERVICE 21 6 2 MAINTENANCE PL...

Страница 3: ...detail the characteristics of the device and the related information The illustrated operating steps are convenient for users Please following this instruction carefully when operating this analyzer 1 1 Introduction of the Analyzer 1 Model number specification FIC Q100N 2 Basic size 280 240 130 length width height unit mm 1 2 Fields of application The Fluorescent Immunoassay Analyzer needs to have...

Страница 4: ...ies after unpacking Note When noticing any missing or damaged accessories please contact Beijing Dianotech Sci Tech Co Ltd and your local sales representatives For contact details please refer to Chapter X of the Operation Instructions 2 1 Standard configuration list of analyzer S N Name Qty Unit 1 Fluorescent immunoassay analyzer 1 Set 2 Power supply adapter 1 Set 3 ID card 1 Piece 4 Quality cont...

Страница 5: ...earance of the analyzer and its accessories is subject to the physical object 1 Front side structure diagram Figure I Front side structure of analyzer 2 Power supply adapter Figure II Power supply adapter Notes This accessory is used in coordination with different testing reagents ...

Страница 6: ...of 86 106 kPa Take care to prevent moisture impact and strong vibration in transportation 3 3 Operating conditions of analyzer 3 3 1Positioning and placing requirements 1 The analyzer should be placed in a stable and horizontal table in a room free from heavy dust direct sunlight and corrosive gas The worktable must be able to bear a weight above 2 5kg 2 There should be no strong vibration sources...

Страница 7: ...ions of analyzer 1 Place the fluorescent immunoassay analyzer on a stable operating table 2 Connect the power supply adapter with the power supply interface of the fluorescent immunoassay analyzer 3 Switch the power supply to the position on to start the analyzer Chapter V Detailed operation steps of analyzer The operation of this test analyzer relies on fingers to operate the touch screen of the ...

Страница 8: ...ject management and historic record operation 5 1 System setup Click the Settings button to enter into the operation interface then process to function setup tools time setup about and factory resetting menu 5 1 1 Function setup Click Function Setup button to enter the following interface Then click icon corresponding to second level items to select the setup items ...

Страница 9: ... an ID number manually 3 Start the built in barcode on scan built in barcode to identify the testing item and batch number information before test off test directly without scanning the built in barcode 5 1 2 Tools Click Tools button to enter the follow interface The main tools include updating program backlight adjustment quality inspection administrator network configuration port configuration a...

Страница 10: ...owing interface You can select network updating or u disk updating Click to copy the updating program package to a U disk and insert it into the USB port behind the analyzer The analyzer obtains the program of a new version from the U disk and updates the system ...

Страница 11: ...t Click the Backlight Adjustment button to enter the following interface Click or to adjust the backlight brightness Click to complete the setup and close the dialogue box III Quality inspection Click the Quality Inspection button to enter the following interface ...

Страница 12: ...ection quality Inspection successful will display on the screen When not passing the quality inspection quality inspection failed will display If the quality inspection fails you have to contact the manufacturer for analyzer adjustment and the testing result is invalid IV Administrator The administrator function is available for professional technicians to adjust the machine only ...

Страница 13: ...val Besides the analyzer provides the following new tools calibration tool gain adjustment save data clear records and about the analyzer 5 1 3 Time setup Click Time Setup button to enter the following interface By clicking Time Setup you can set up time Click the field to be modified and ...

Страница 14: ...n click Time Setup to confirm the modification Clicking Cancel to cancel time modification 5 1 4 About the analyzer Click About to enter the following interface which displays the model number and version information of the fluorescent immunoassay analyzer ...

Страница 15: ... can clear all data to resume the ex factory setup After resuming the ex factory setup the machine restarts automatically 5 2 Project management Click Project Management to enter the following interface Clicking the corresponding item to select the information appropriate for current test ...

Страница 16: ...es to ID card import 1 Each item number and batch number have a unique corresponding examination information data Setting again will overwrite the original setup information 2 When inserting the ID card to the analyzer to scan for the matched barcode and complete one test it will add the current ID card inspecting item to the list 5 3 Single sample testing Step 1 Install the analyzer properly and ...

Страница 17: ...5 2 Project Management when it is necessary to change Step 5 After obtaining the confirmed information begin to test Click and the analyzer start the motor to scan the card and begin to test If clicking the analyzer begin to count the incubation time The test will start only after incubation is finished ...

Страница 18: ... After test is finished the result will display on the screen Step 6 Check the Result When the test is complete the result will display on the screen By clicking the detailed test result could be checked ...

Страница 19: ...17 5 4 Batch testing Begin to incubate during incubation inserting a test strip testing pull out the test strip Click Batch Testing to switch to the following interface ...

Страница 20: ...to complete adding the IDs of all samples to be tested Step 2 Begin to incubate Add samples to the test strips Immediately after adding samples click Begin to Incubate The testing state changes and incubation time countdown begins ...

Страница 21: ...o test When incubation timing is finished put in the test strip Click Start Test to begin testing When the test is finished the test result will display on the screen The test is completed when no more results displayed ...

Страница 22: ...se the test records Step 4 Repeat the above operations to test the next strip 5 5 Historic records Click the Historic Records button on the main menu to switch to the following interface The operation records are provided ...

Страница 23: ...ple ID item number and enter the retrieval field By clicking the Search button you can find the corresponded record Chapter VI Maintenance and Service 6 1 Maintenance What you have to do is only to keep the external part of the fluorescent immunoassay analyzer clean External cleaning and maintenance method Clean the outer surface of the analyzer with wet cloth and 70 ethanol Do not use any strong ...

Страница 24: ...perate or inconvenient to disconnect the device 6 3 2 Except the test cards provided by the manufacturer do not place anything else into the test strip holder 6 3 3 If any test samples has potential infections please use protective gloves or take other protective measures to avoid skin contacting with the tested samples 6 3 4 Please dispose all used reagents and strips according to the Medical Was...

Страница 25: ...oading of the software installation package in windows XP windows7 Chapter VII Troubleshooting Guide Steps dealing with device crash When there is no response to clicking the analyzer interface buttons or analyzer press keys device crash occurs Please restart the analyzer by following the steps below 1 Switch off the power 2 Plug and unplug the power cord again 3 Switch on the power Notes to commo...

Страница 26: ...lyzer has any trouble or the upper computer software operates abnormally please call 400 0536917 for consultation We will provide technical supports online to help customers for troubleshooting If any analyzer needs repairing please send the analyzer back for repairing according to the relevant requirements Normally the repair period is not beyond one week If any analyzer does have to be scrapped ...

Страница 27: ... Co Ltd Registered address No 1 Zhongli Technology Park Shangdi 3 Road Haidian District Beijing Production address No 1 Zhongli Technology Park Shangdi 3 Road Haidian District Beijing Medical apparatus registration certificate number Production license number Contact 400 0536917 Fax 010 82816022 ...