V1.0K3 PULSE OXIMETER INSTRUCTION MANUAL
18
with “
×
”.
Note: Please caution the alarm off item, it may cause damages to users.
ALARM PRIORITY
Alarms of the oximeter include technical and physiological alarms. All the three priorities
are divided by built-in module and can not be changed by users.
High priority:
indicates the patient is in the very dangerous situation.
Medium priority:
indicates the warnings should be paid attention to.
Low priority:
indicates the technical alarm caused by the device itself.
Assignment of priority:
High
Medium
Low
Paramter
SpO
2
PR
Value
Red Yellow
Alarm lamp
Flashing with yellow
Flashing with yellow
Lamp
Frequency
1.5Hz 0.5Hz
Audiblesound
Di- Di – Di ----- Di - Di
Di - Di - Di
Di
Alarm cycle
3 s
5 s
20 s
Alarm info
SpO
2
too high/low
PR too high/low
Probe off/Finger out
AUDIBLE ALARM INHIBITION:
Short press the
button to silence the audible alarm for 120 seconds, the audible
alarm indicator will be displayed as
, together with the countdown from 120s to 0s,
short press it again, you can cancel alarm inhibition.
Warnings!
When
an alarm occurs, check patients’ conditions immediately.
z
Check which parameter is alarming or which alarm is going on.
z
Check
patient’s
condition.
z
Search for the source of alarm.
z
Make the alarm mute if necessary.
6.2 Data Management
From the main menu screen, select and enter the “Data Manage” screen, refer to Fig.6-2.
The unit can store 72-hour records. If the storage is full you should delete the old records