– 16 –
behr S 5
249101
Explanation:
Prog.Number 01
You can modify the settings of pro-
gram no. 01
REAC.TIME
Delay time in seconds between
the end of NaOH addition and the
start of distillation. Possible entry
values range from 0 sec to 999
sec. The factory default setting
is 0 sec. For the distillation in the
Kjeldahl nitrogen determination,
do not change the preset default
of 000.
POWER
Steam generator power. Possible
entry values range from 30% to
100%. The factory preset default
value is 80%. You can change the
electrical power for steam gene-
ration and thereby the intensity of
steam generation.
The steam generator is swit-
ched on and off on by a pressure
sensor. The on/off set points of
the pressure sensor cannot be
adjusted by the user. It is only
possible for the user to adjust the
steam generation power.
DRAINSAMPLE
Run-time in seconds for the pump
draining the sample waste after
distillation. Possible entry values
range from 0 sec to 999 sec. The
factory default setting is 30 sec.
<<==
back
==>>
further
If you wish to change any of the values, select the
desired parameter by turning the operating control knob.
Press on the control knob to access the numerical entry
value. Then dial the desired value by turning the oper-
ating control knob and enter it in memory storage by
pressing the knob again.
Once all entry values have been entered according to
your wishes, select
==>>
and press the operating con-
trol knob to bring up the following program menu:
DrainTitr
010
H3BO3
3.0
AutoEndPt
Off
==>>
<<==
Prog.Number
01
Explanation:
AutoEndPt
Automatic endpoint determination.
Before destillation, the S 5 deter-
mines the pH of the boric acid;
afterwards it titrates back to this
pH. If AutoEndPt is Off, titration
goes to a fixed endpoint. Not all
titrators allow automatic endpoint
determination!
DrainTitr
Run-time in seconds for the pump
draining the distillate after titration.
Possible entry values range from
0 sec to 999 sec. The factory
default setting is 30 sec.
H3BO3
Delivery time for boric acid into the
titration vessel in seconds.
Possible entry values range from
0.0 sec to 99.9 sec. The factory
default setting is 3.0 sec.
Delivery of the pump approx.
14 ml/sec.
70 ml/5 sec
840 ml/min
<<==
back
==>>
further
If you wish to change any of the values, select the
desired parameter by turning the operating control knob.
Press on the control knob to access the numerical entry
value. Then dial the desired value by turning the oper-
ating control knob and enter it in memory storage by
pressing the knob again.
Once all entry values have been entered according to
your wishes, select
==>>
and press the operating con-
trol knob to bring up the following program menu:
Meas.Time
020
Stirring
40%
SetPoint
4.6
<<==
Prog.Number
01
Explanation:
SetPoint
Endpoint preset for the titra-
tion. Titration will go to this pH if
AutoEndPt is off.
Not all titrators allow endpoint
setting from the S 5. With some
titrators the endpoint must be set
on the titrator itself.
Meas.Time
Time in seconds for the endpoint
determination if AutoEndPt is on
and the titrator allows automatic
endpoint determination. The fac-
tory default setting is 10 sec.
Distilling