17
MAINTENANCE AND REPAIR
If the meter fails to operate, check battery, test leads, other accessories, and
replace as necessary.
Double check the followings:
1. Replace the fuse or battery if the meter does not work.
2. Review the operating instructions for possible mistakes in operating
procedure.
Quick check on 0.5A FUSE:
Step 1:
Turn the rotary switch to
Ω
function.
Step 2:
short-circuit
E
/V/
Ω
/Hz terminal and mA/μA terminal.
Resistance reading
≤
1M
Ω
: the fuse is OK
Resistance reading “OL”: the fuse is open. Replace the fuse as specified.
Quick check on 10A FUSE:
Step 1:
Turn the rotary switch to
Ω
function.
Step 2:
short-circuit
E
/V/
Ω
/Hz terminal and A terminal.
Resistance reading
≤
0.5
Ω
: the fuse is OK.
Resistance reading “OL”: the fuse is open. Replace the fuse as specified.
Содержание 0095969800341
Страница 2: ......
Страница 25: ...19 F1 9V ba tte ry F2 ...
Страница 26: ......
Страница 50: ......
Страница 74: ......
Страница 98: ......
Страница 121: ...19 F1 9V ba tte ry F2 ...
Страница 122: ......
Страница 146: ......
Страница 169: ...19 F1 9V ba tte ry F2 ...
Страница 170: ......
Страница 193: ...19 F1 9V ba tte ry F2 ...
Страница 194: ......
Страница 195: ...AM 535 EUR Digitalt multimeter Brukerhåndbok 6 2015 6005740 A 2015 Amprobe Test Tools Med enerett Trykt i Kina Norsk ...
Страница 217: ...19 F1 9V ba tte ry F2 ...
Страница 218: ......
Страница 241: ...19 F1 9V ba tte ry F2 ...
Страница 242: ...20 ...
Страница 266: ......
Страница 289: ...19 10 A indgangsterminal F2 sikring 11 A H 1000 V hurtig sikring 10 38 mm F1 9V ba tte ry F2 ...
Страница 290: ......