Open the Bag Clamp and press the next arrow to 

continue.

An hourglass will display onscreen as the device

will automatically calibrate, drawing the gas 

from the CaliBag®.

A successful calibration is shown by a green tick. 

Press Save  

  to complete the calibration.

An unsuccessful calibration is shown by a red cross. 

 

If an unsuccessful calibration is shown by a red 

cross, contact MGC Diagnostics Corporation or the 

local distributor.

Ensure both Bag Clamps are closed before 

removing the CaliBag® and remove all adapters 

from the FeNObreath™️.

Remove the Bung from the FeNObreath™️.

11

10

9

12

13

14

Please ensure the FeNObreath™ monitor is 

switched off before removing the back cover.

Using the screwdriver provided with the  

FeNObreath™, unscrew the screw on the back of 

the monitor.

Remove the back cover by sliding 

 

and lifting it off.

To remove the scrubber, first press down on 

 

clip at the top left-hand side of the monitor.

Slide out the scrubber then lift to remove.

Remove all packaging from the new NO 

 Scrubber and insert it into the FeNObreath™️.

Re-attach the back cover of the

 

FeNObreath™️ monitor.

To secure the FeNObreath™ back cover, use the 

screwdriver to resecure the screw into place.

How to replace the NO Scrubber

4

5

6

7

1

2

3

8

(

CAUTION:

 NO scrubber contains potassium permanganate and 

should not be tampered with and should not be exposed to 

skin.)

Содержание FeNObreath CaliBag

Страница 1: ... to Calibrate the FeNObreath with the CaliBag Page 4 How to replace the NO Scrubber Page 6 Disposal Information Page 7 CaliBag b a c d e f g a CaliAdapter x1 b Bung x1 c CaliBag x1 d Bag Adapter e Blue Plugs x2 f Bag Clamps x2 g NO Scrubber x1 THIS KIT CONTAINS FeNObreath CaliBag ...

Страница 2: ... 25 C or 77 F a red thermometer will be shown on screen Move the FeNObreath to a cooler area and try again later Safety Information CAUTION NO scrubber contains potassium permanganate and should be disposed of as hazardous waste in accordance with local waste disposal regulations CAUTION NO scrubber contains potassium permanganate and should not be tampered with and should not be exposed to skin N...

Страница 3: ...Obreath 11 10 9 12 13 14 Please ensure the FeNObreath monitor is switched off before removing the back cover Using the screwdriver provided with the FeNObreath unscrew the screw on the back of the monitor Remove the back cover by sliding and lifting it off To remove the scrubber first press down on clip at the top left hand side of the monitor Slide out the scrubber then lift to remove Remove all ...

Страница 4: ... scrubber contains potassium permanganate and should be disposed of as hazardous waste in accordance with local waste disposal guidelines Packaging The CaliBag Annual Service Kit comes in carboard packaging this is widely recycled Please follow local waste guidelines on how to recycle or dispose of carboard The NO Scrubber comes in a plastic bag which is made of polyethylene this is also recyclabl...

Отзывы: