There is a separate and distinct temporary 25 second hold time programmed into
the Control Unit. When weight is on the Sensormat
®
(patient on the mat in bed
or on the chair) and the Control Unit is
MON
itoring (green light), the caregiver
may press the
RESET
button to put the system into temporary
HOLD
status
(blinking yellow light) for servicing the patient. The unit will remain in this
temporary
HOLD
status as long as weight stays on the Sensormat
®
for up to 25
seconds, and then will resume
MON
itoring. However, if any time during the 25
seconds weight is removed from the Sensormat
®
, this immediately puts the unit
into permanent
HOLD
(steady yellow light). Once on permanent
HOLD
, the
unit will begin the preselected
HOLD
delay countdown (e.g. 5 seconds) to
MON
itor only when weight is returned to the Sensormat
®
.
Note that in any
HOLD
mode, monitoring may be started immediately (no
delay countdown) by pressing
RESET
, as long as weight is on the Sensormat
®
.
Always assure the green MON
itor
light is on before leaving the patient.
The caregiver may adjust the patient/resident in bed, do local therapy, etc. dur-
ing the (25 second) temporary
HOLD
period without setting off the alarm. If
the Control Unit begins to again monitor and it is taking longer to care for or
remove the patient/resident from the bed, press
RESET
again to renew the 25
second temporary
HOLD
time. If the care giver has completed the task and the
control unit is still in the temporary (blinking)
HOLD
status, press
RESET
to
immediately begin
MON
itoring before leaving the room.
MON
itor - The green
MON
indicator is illuminated when the control unit is
monitoring the patient. If the patient’s weight is off the Sensormat
®
for the entire
preselected delay, i.e. 3 seconds, the alarm will sound. If
RESET
is pressed
while the
MON
indicator is active, the control unit will return to the
HOLD
mode. Please review the
HOLD
function above. A flashing
MON
itor indicator
indicates that the unit is actively monitoring and that weight is off the
Sensormat
®
, but the preselected alarm delay has not yet expired.
ALARM
- The red
ALARM
indicator is illuminated when the control unit is
alarming. The control unit will exit the alarm mode and enter the
HOLD
mode
when the
RESET
button is pressed. The alarm delay time can be set within a
range of 1 to 9 seconds. We recommend 3 seconds. The time delay should be set
just long enough to allow patient movement in bed without generating false
alarms.
The nurse call system must be physically reset in the patient’s room after reset-
ting the Control Unit in the event of an alarm.