![Beckman Coulter UniCel DxC Synchron 600 Скачать руководство пользователя страница 48](http://html.mh-extra.com/html/beckman-coulter/unicel-dxc-synchron-600/unicel-dxc-synchron-600_instructions-for-use-manual_2719768048.webp)
Cuvette Reaction System
Photometer Assembly
System Description
UniCel DxC Systems Instructions For Use A13914
Page 2-22
October 2005
Figure 2.14
Reaction Carousel Area (Typical – Cover Removed)
Photometer Assembly
Attached to the reaction carousel support frame is the Photometer assembly. This
consists of a xenon pulse lamp, a discrete 10-position silicon-diode detector array, a
monochromator housing unit, and associated electronic circuitry. (Refer to
As each cuvette passes through this optics station during a spin cycle, the xenon lamp
is flashed and the resulting light beam travels through the opposing sides of the square
cuvette.
LPIA (Large Particle Immuno Assay) or NIPIA (Near-Infrared Particle Immuno Assay) Module
(optional)
The optional LPIA module uses a photometric detection system for large particle
immuno assays. This assembly is attached to the reaction carousel support frame to the
left of the photometer assembly, near the sample carousel. (Refer to
has two printed circuit boards, an LED (light-emitting diode) and a photodetector.
Communication with the system software is through fiber optic cables.
1. LPIA module
2. Reaction carousel
3. Photometer
A015909P.EPS
2
1
3