Reagent Load
Rgts/Cal Screen Overview
Reagent Load/Calibration
UniCel DxC Systems Instructions For Use A13914
Page 5-2
October 2005
Rgts/Cal Screen Overview
If the Rgts/Cal icon is red or yellow, select it and look for chemistries that are
highlighted red or yellow. (Refer to
.) Red highlighted chemistries will not
run. Yellow highlighted chemistries will run, but check the reagent volume before
starting. Load new reagent if the reagent volume is insufficient to run pending tests.
Figure 5.1
Reagent Load/Calibration Screen
Load requests can be made simultaneously or independently for the MC and CC sides
of the instrument, and can be made while the system status is in Standby or Running.
The system automatically senses the level of reagent in all the cartridges. Reagent
status messages are listed in
Table 5.1
Reagent Status Messages
Message
Explanation
Load Requested
Reagent position has been requested for reagent load but
the system is running.
Parameters Required
Parameters for a reagent cartridge are missing and were not
loaded when the cartridge was loaded.
Reagent Ok
No error flags present.
Reagent Expired
Reagent has exceeded shelf life stability date.
Days Exceeded
Reagent has been on system longer than acceptable.
0 Tests Available
Reagent has been used completely; reagent cartridge is
empty.
Level Pending
Reagent cartridge has been loaded and is waiting to be
level-sensed.
(1 of 2)
E016483S.EPS