A13914AF
6-4
Sample Programming and Processing
Identify Samples
Verify Sample Carousel Status: Main Screen
The status of racks currently loaded on the sample carousel can be viewed on the
Main
screen. The
status of each sample is designated by an icon, described by the legend at the top of the screen.
Refer to CHAPTER 1,
System Description
for additional information on Sample Carousel
status.
Identify Samples
Introduction
Samples may be identified by a Sample ID or by a rack/cup position or both.
Rack and Position Mode and Bar Code Mode setup options affect where the cursor starts within
sample programming and which fields are required to save sample programs. If rack and position
priority is chosen during system setup, the cursor begins at the
Rack
field. If Bar Code Priority is
chosen, the cursor appears at the
Sample ID
field, skipping the rack and position assignment. In
both cases, either a rack and position or a Sample ID assignment is required, but both can be
entered.
Refer to CHAPTER 3,
System Setup
for additional information on setting up sample
identification features.
If a rack has a readable bar code and a manual rack position assignment for the same sample
position, the system verifies that the bar code and the manual assignment match. If there is a
discrepancy, the rack is offloaded and the operator notified with a message. If the bar code and the
manual assignment agree, the rack remains on the carousel and is processed.
Table 6.1
Sample Status Designations
Sample Status
Meaning
In Progress
Sample has been identified and is currently being sampled for programmed
tests.
Incomplete
Sample has completed all tests possible and has some tests which are still
pending.
Complete
All tests have been completed.
Reserved for Cal
Rack/Position assigned to a calibrator.
Sample Required
Sample is programmed but not yet run by the system.
Rerun
Sample has been requested for rerunning.
Removed
Sample has been unloaded but some tests are still being processed.
Manual Assign
Sample ID has been manually assigned to a Rack/Position.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......