19
SW 41 Ti Rotor
➠
NOTE
Do not immerse the rotor body in water, since
the hanger mechanism is difficult to dry and
can rust.
3. Rinse the cleaned rotor and components with distilled water.
4. Air-dry the buckets upside down.
Do not use acetone to dry
the rotor
.
Clean metal threads frequently to prevent buildup of residues and
ensure adequate closure. Use a brush and concentrated Solution 555.
Rinse and dry thoroughly, then lubricate lightly but evenly with
Spinkote to coat all threads.
DECONTAMINATION
If the rotor or other components are contaminated with toxic,
radioactive, or pathogenic materials, follow appropriate decontamina-
tion procedures as outlined by your laboratory safety officer. Refer to
Appendix A in
Rotors and Tubes
to select solutions that will not
damage the rotor and accessory materials.
STERILIZATION AND DISINFECTION
• The rotor and all rotor components can be autoclaved at 121°C for
up to an hour. Remove the lids from the rotor buckets and place the
rotor, buckets, lid, and spacers in the autoclave upside down.
• Ethanol (70%)
*
or hydrogen peroxide (6%) may be used on all
rotor components, including those made of plastic. Bleach (sodium
hypochlorite) may be used, but may cause discoloration of
anodized surfaces. Use the minimum immersion time for each
solution, per laboratory standards.
*
Flammability hazard. Do not use in or near operating ultracentrifuges.