4-5
Troubleshooting and Maintenance
Figure 4-3. Interior View of Instrument
7. After retrieving the sample, close the chamber door, then replace
the front panel on the instrument. Insert the tabs at the bottom of
the front panel onto the lip at the base of the instrument, being
sure to align the front panel with the side panels (see Figure 4-4).
8. Lift the front edge of the top panel a few inches and insert the
upper edge of the front panel under it. Push back gently to engage
the tabs, then down to activate the latch.
9. Replace the control head by sliding it along the small ridges pro-
vided, making sure to clear the emergency POWER OFF switch in
front. When the control head cover is correctly positioned, push it
until the latches snap shut and the cover is aligned with the rest of
the instrument panels.
WARNING
!
Do not attempt to run the instrument before
returning the front panel to its correct
position.
Spring-loaded
Interlock Pin
Vacuum
Solenoid
Screw
Fan housing
Door
Lock
System