773232AH
4-17
Specifications
Software Specifications
4
There are three additional keywords follow the TEXT segment: @EOFVERSION, REFERENCE and
@Acquisition Protocol Offset. These keywords are delimited by the “!” character and are not
contained within the offsets specified for the TEXT segment. These keywords are reserved for
Beckman Coulter’s use only.
DATA Segment
The DATA segment contains the 20 bit, uncompensated raw data produced by the instrument. Each
event contains data for the parameters specified in the TEXT segment. The number of parameters
per event is specified by the value of the $PAR keyword. The order of the parameters within each
event is specified by the parameter indices used in the TEXT segment. The parameter data has a 20
bit range and it occupies 32 bits (4 bytes) in the file. The parameters are aligned to the byte
boundary. The data for each parameter is right justified with 12 bits of padding in front in little
endian order, meaning that the least significant byte goes first.
Following, there is the ordered list of parameters as reported by the cytometer. Up to a maximum
of 16 parameters from this list can be collected. Parameter indices are assigned incrementally
starting from 1 following the order listed.
FS Peak, FS Int, FS TOF, SS Peak, SS Int, SS TOF, FL1 Peak, FL1 Int, FL1 TOF, FL2 Peak, …, FL9 TOF, FL10
Peak, FL10 Int, FL10 TOF, TIME
Database Structure
The information in this section specifies the read-only access to Navios software for 3rd party LIS
vendors. The information in this section includes,
•
•
•
Report Generator Database Schema
NOTE
Beckman Coulter does not warranty data after 3rd party middleware has retrieved data from the
Navios report database.
Firewall Setup
Use these instructions to setup the Windows Firewall to allow SQL Server application access.
@TESTNAME
Protocol name.
@TUBENO
Tube number read from the sample loader.
@VIOLETLASERPOWER_END
Violet laser power at end of acquisition.
@VIOLETLASERPOWER_START
Violet laser power at start of acquisition
@VIOLETLASERSHUTTER
Violet laser shutter status: “Open” or “Closed”.
@VIOLETTARGETPOWER
Violet laser target power.
@Y2KDATE
Date in format YYYYMMDD.
Table 4.2
FCS 3.0 Keyword Information
FCS 3.0 Keyword
Key Value
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......