PN A51387A
B
4-15
Selecting Analysis Settings: SOM and SOP
Using a Standard Operating Method (SOM)
4
Analyzer as you edit settings. You can use the
(
Threshold
)
and
(
Current & Gain
)
buttons on the
Current & Gain tab to measure noise level and resistance using the currently installed aperture tube.
Standard Operating Method: Pulse to Size Settings
Use Step 6 of the SOM Wizard (page
) or the Pulse to Size Settings tab of the Edit the SOM window
(page
) to enter the number of bins (channels), bin spacing, and other information that will
determine the range and resolution of analysis results. After running the analysis, you can change
these settings by selecting
Calculate > Convert Pulses to Size
on the Run Menu bar. The aperture
information on this screen changes accordingly, depending on the size of the aperture tube you are
using.