Error Indicators
Express 3 error conditions are signaled by a flashing red Error or Service indicator on the
front panel.
The cause of the error condition can be identified by the combination of green cycle time
indicators that are also flashing.
Error Codes
Table 2.2
Cycle Time Indicator Combination
Problem
Solution
2 min light and red indicator
flashing
Rotor failed to reach full
speed in 30 seconds
Confirm no obstruction present in
bowl. Contact customer service if
error continues to occur.
3 min light and red indicator
flashing
Cover not latched
Confirm cover is fully seated in
closed position. Contact customer
service if error continues to occur.
5 min light and red indicator
flashing
Unable to maintain rotor
speed
Confirm no obstruction present in
bowl. Contact customer service if
error continues to occur.