Installation
Installation Requirements
It is recommended that a surge protector be used to ensure correct operation in the presence of
electrical transients and noise. At least three 3-wire service outlets are needed for the instrument
with its computer, monitor and printer.
Figure 2.1
Electrical Specification Labels
Temperature and Humidity Requirements
Temperature: 10 to 35° C
Relative Humidity: Zero to 85% without condensation
Sample Modules Specific Requirements
Suspension Fluid for Aqueous and Universal Liquid Modules
IMPORTANT
Ensure that the material being sized and the suspension fluid do not violate any hazard waste
or clean water discharge regulations.
The Aqueous Liquid Module requires clean, bubble-free liquid as suspension fluid. Regular tap water
that is filtered and degassed can be used. Beckman Coulter recommends that you use a 0.2 µm or
smaller pre-filter.
Table 2.1
Table 4.1 LS 13 320 Power Requirements
Power Category
Requirement
Input Voltage (V)
100, 120, 220, 240
Input Frequency (Hz)
50/60
Power Consumption (W)
720 Maximum
Содержание LS 13 320
Страница 4: ...PN B05577AC iv Revision History ...
Страница 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Страница 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Страница 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Страница 58: ...PN B05577AC 2 22 Installation Making Measurements Figure 2 19 Reference Background File Selection ...
Страница 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Страница 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Страница 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Страница 105: ...PN B05577AC 3 47 LS 13 320 Software High Security 3 Figure 3 5 User Privileges Dialog High Security ...
Страница 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Страница 118: ...PN B05577AC 5 2 Data File Menus RunFile Menu 3 Select the file you wish to overlay with the current opened file 4 Open ...
Страница 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Страница 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Страница 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Страница 238: ...PN B05577AC C 10 Optical Models Statistics ...
Страница 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 253: ......