The critical speed range is the range of speeds over which the rotor shifts so as to rotate about
its center of mass. Passing through the critical speed range is characterized by some vibration.
†
Relative Centrifugal Field (RCF) is the ratio of the centrifugal acceleration at a specified
radius and speed (
r
ω
2
) to the standard acceleration of gravity (
g
) according to the following
formula:
where
r
is the radius in millimeters,
ω
is the angular velocity in radians per second (2
π
RPM /
60), and
g
is the standard acceleration of gravity (9807 mm/s
2
). After substitution:
RCF
r
ω
2
g
---------
=
RCF
1.12
r
RPM
1000
------------
2
=
4
J-Lite JLA-16.250 FIXED ANGLE ROTOR
SPECIFICATIONS
Maximum speed (Avanti J series centrifuges) . . . . . . . . . . . . . . 16 000 rpm
Maximum speed (J2 series centrifuges) . . . . . . . . . . . . . . . . . . . 14 000 rpm
Critical speed range* . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 600 to 800 rpm
Maximum solution density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2 g/mL
Relative Centrifugal Field
†
at maximum speed
at
r
max
(134 mm) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 400
×
g
at
r
av
(90 mm) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 800
×
g
at
r
min
(46 mm) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 200
×
g
k
factor at max. speed in Avanti J series centrifuges. . . . . . . . . . . . . . . 1090
k
factor at max. speed in J2 series centrifuges . . . . . . . . . . . . . . . . . . . 1400
Conditions requiring speed reduction . . . . . . . . . . . . . . . . see
RUN SPEEDS
Maximum allowable imbalance of opposing loads. . . . . . . . . . . . . 10 grams
Number of bottle cavities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Available bottles and tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . see Table 3
Tube angle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25°
Nominal tube capacity of largest tube. . . . . . . . . . . . . . . . . . . . . . . . 250 mL
Nominal rotor capacity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1500 mL
Rotor weight, empty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.3 kg (22.7 lb)
Rotor weight, maximum sample mass . . . . . . . . . . . . . . . . 13.2 kg (29.1 lb)
Approximate acceleration time to maximum speed
(rotor fully loaded)
in an Avanti J centrifuge (to 16 000 rpm) . . . . . . . . . . . . . . . . 3
1
/
2
min
in a J2 series centrifuge (to 14 000 rpm) . . . . . . . . . . . . . . . . . 4
1
/
2
min
Approximate deceleration time from maximum speed
(rotor fully loaded)
in an Avanti J centrifuge (from 16 000 rpm) . . . . . . . . . . . . . . 2
3
/
4
min
in a J2 series centrifuge (from 14 000 rpm) . . . . . . . . . . . . . . . . . 4 min
Rotor material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . aluminum
Rotor entry code for Avanti J-E centrifuges . . . . . . . . . . . . . . . . . . . . . 16.25
Rotor entry code for
microprocessor-controlled J2 series centrifuges . . . . . . . . . . . . . . . . 14
25°
r
min
Axis of Rotation
r
av
r
max