Routine Maintenance
354
CEQ™ 8000 Genetic Analysis System
Figure 218:Cleaning the Detection Window
10. With a new water-moistened swab, repeat wiping on the other side of the window.
11. With a dry swab, gently wipe the windows to remove excess water, again repeating
on the backside with a new, dry swab.
12. Blow compressed gas on the windows to remove all excess water.
Care must be taken not to invert the compressed gas bottle,
otherwise propellant will contaminate the capillary windows.
13. If dried gel or other debris remains on the window, repeat this procedure.
14. From the
Remove
Capillary
dialog box (Figure 219), select the
Clean
Capillaries
radio button and click
OK
.
900501e.AI
900502e.AI
900501e.AI
900502e.AI
Detection Window
Stroke in one
direction
ONLY
!
Array Fitting
CAUTION
Содержание CEQ 8000
Страница 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Страница 98: ...84 CEQ 8000 Genetic Analysis System...
Страница 110: ...96 CEQ 8000 Genetic Analysis System...
Страница 120: ...106 CEQ 8000 Genetic Analysis System...
Страница 128: ...114 CEQ 8000 Genetic Analysis System...
Страница 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Страница 154: ...140 CEQ 8000 Genetic Analysis System...
Страница 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Страница 208: ...Sequence Analysis Procedures 194 CEQ 8000 Genetic Analysis System Figure 119 Sequence Results Report...
Страница 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Страница 303: ...User s Guide 289 More information on these and other display modifications can be found in the Online Help...
Страница 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Страница 329: ...Exporting Database Items User s Guide 315 Figure 180...
Страница 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Страница 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...