Run Procedures
144
CEQ™ 8000 Genetic Analysis System
The system will measure the gel volume accurately and will determine whether
or not to proceed with the run.
4. If the actual volume (re-estimated by the system) is
greater
than 1.5 mL more than
the total volume needed to run the plate, the system will automatically start the
plate.
5. If the actual volume (re-estimated by the system) is
less
than 1.0 mL more than the
total volume to run the plate, you will be presented with the dialog box in
Figure 85.
Figure 85: Gel Usage Confirmation Dialog
6. Click
Accept
to proceed with the run. If the run is selected to continue, specify the
sample sets to run. Select
Replace Gel Cartridge
to replace the gel cartridge
before the run.
Gel Volume Notification
A dialog box will warn you if you attempt any operation requiring gel when there is
not enough gel in the cartridge to perform the operation.
You will be presented with the warning message in Figure 86 if you attempt to
perform a capillary gel fill when the cartridge does not have enough gel.
Figure 86: Gel Warning Dialog
Содержание CEQ 8000
Страница 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Страница 98: ...84 CEQ 8000 Genetic Analysis System...
Страница 110: ...96 CEQ 8000 Genetic Analysis System...
Страница 120: ...106 CEQ 8000 Genetic Analysis System...
Страница 128: ...114 CEQ 8000 Genetic Analysis System...
Страница 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Страница 154: ...140 CEQ 8000 Genetic Analysis System...
Страница 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Страница 208: ...Sequence Analysis Procedures 194 CEQ 8000 Genetic Analysis System Figure 119 Sequence Results Report...
Страница 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Страница 303: ...User s Guide 289 More information on these and other display modifications can be found in the Online Help...
Страница 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Страница 329: ...Exporting Database Items User s Guide 315 Figure 180...
Страница 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Страница 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...