Preparing Other Integrated Devices
Electrical Component Requirements
4
Electrical Component Requirements
Make sure the proper electrical supply is available at the location for the Biomek NX
P
instrument.
Table 4.1, Electrical Components and Requirements
defines the electrical requirements for specific
integrated devices.
NOTE
All AC devices are at 50/60 Hz.
Table 4.1
Electrical Components and Requirements
Electrical Component
Electrical Requirements
Bar Code Reader
110-220 VAC @ 1 amp max.
Biomek NX
P
instrument
100-240 VAC @ 4 amps max.
Circulating Reservoir/Tip Box ALP
115 VAC @ 1.5 amps
220 VAC @ 0.8 amps
Cytomat 2
100-230 VAC @ 2.6 amps
Cytomat 6001 Incubator
100-120 VAC @ 9 amps
230 VAC @ 4.4 amps
Cytomat ALP
100-240 VAC @ 1.2 amps
Cytomat Microplate Hotel
100-230 VAC @ 0.7 amps
I/O Box
100-240 VAC @ 6.3 amps
WINDOWS 7
Automation Controller
Monitor
100-127 VAC @ 5 amps max.
200-240 VAC @ 3 amps max.
100-240 VAC @ 1.8 amps max.
WINDOWS 10
Automation Controller
Monitor
100-240VAC @ 2.5 amps
100-240VAC @ 1 amp