Section 3: Service Procedures
When the software detects the chamber temperature (thermistor) exceeds 50°C, or the software reads
the can over temp sensor it will generate an Over temp diagnostic and shutdown the instrument by
coasting. There is no minimum chamber temperature requirement or diagnostic.
Diag #
Description
Action taken
When issued
Display
T1
Max temperature exceeded
Shutdown no brake,
door locked for 45 min
When thermistor reading
exceeds 50C
T-1 Temp, door locks for
45 minutes, call Service
T2
Chamber overtemp
Shutdown no brake,
lock door 45 minutes
HW sends signal to
control board
T-2 Temp, door locks for
45 minutes, call Service
T3
Motor over temp
Shutdown, no brake
HW sends signal to
control board
T-3 Temp, door locks for
45 minutes, see manual
Refrigeration Diagnostics
When the software detects the set temperature is not controlled within a temperature range an error will
be displayed
Diag #
Description
Action taken
When issued
Display
C1
Temp exceeds 5C above set
No action
If reached temp and 4C
over set
C-1 Rotor temp exceeds
4C above set
C2
Temp exceeds 8C above set
Shutdown no brake,
If reached temp and 8C
above set
C-2Rotor temp exceeds 8C
above set, call Service
C3
Temperature not controlled
Shutdown with brake
Not trending towards
set temp
C-3 Temp Error,
call Service
C5
Temperature bad reading
Shutdown with brake
Software read an
ambiguous temperature
reading
C-5 Temp Error,
call Service
Drive Diagnostics
Diag #
Description
Action taken
When issued
Display
D1
Fault not cleared
Shutdown no brake
Driver board issues fault;
SW cannot clear
D-1 Drive error,
call Service
D2
Fault limit exceeded Shutdown
no
brake More
than
2 driver faults
occurred during run
D2-Drive Error, call
Service
D4
Bus voltage limit exceeded
Shutdown, no brake
SW detects bus voltage
over maximum allowed
voltage 435 Volt
D-4 Drive error, call
Service
D5
Driver not powered
Shutdown, no brake
Drive power lost during
run
D-5 Drive Error, call
Service
D6
Driver over temperature
Shutdown, no brake
Signal from driver board
indicates an over temp
D-6 Drive Error, call
Service
D7
Driver not enabled
Shutdown, no brake
Driver not enabled after
turning On
D-7 Drive Error, call
Service
D8
Driver not established
Shutdown, no brake
Driver not established
after turning Off
D-8 Drive Error, call
Service
D9
Driver not connected
Shutdown, no brake
Driver board not
connected
D-9 Drive Error, call
Service
Speed Diagnostics
3-18
Allegra X-12/R, X-15R Service Manual 07/07
Содержание ALLEGRA X-12/R
Страница 1: ...Part Number 679060 Rev AA ALLEGRA X 12 R X 15R Service Manual...
Страница 13: ...SECTION 1 INTRODUCTION 1 1 ALLEGRA X 12 R X 15R PERFORMANCE SPECIFICATION 1 1 1 2 GENERAL DESCRIPTION 1 3...
Страница 17: ...SECTION 5 PARTS 5 1 REPLACEMENT PARTS LIST 5 1 5 2 TABULATION DRAWING X 12 R 5 5 TABULATION DRAWING X 15R 5 13...
Страница 18: ...SECTION 6 ACCESSORIES MISCELLANEOUS NO MATERIAL AVAILABLE AT THIS TIME...
Страница 19: ...SECTION 7 APPLICATIONS NO MATERIAL AVAILABLE AT THIS TIME...
Страница 20: ...SECTION 8 SERVICE MEMOS ISSUED DURING TRAINING CLASS...
Страница 25: ...Section 1 Introduction Allegra X 12 R X 15R Service Manual 07 04 1 5...
Страница 43: ...Section 3 Service Procedures Figure 3 1 Allegra X 12R 3 10 Allegra X 12 R X 15R Service Manual 07 07...
Страница 44: ...Section 3 Service Procedures Allegra X 12 R X 15R Service Manual 07 07 3 11...
Страница 45: ...Section 3 Service Procedures Allegra X 15R 3 12 Allegra X 12 R X 15R Service Manual 07 07...
Страница 60: ......
Страница 61: ......
Страница 62: ......
Страница 63: ......
Страница 64: ......
Страница 65: ......
Страница 66: ......
Страница 67: ......
Страница 68: ......
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......