Access from the user program
KL3356
40
Version: 2.3.0
Bit
Name
Description
CB2.3 RegLockReq
0
bin
Register lock not active:
• The KL3356 can update the values in registers
,
and
can be shown by setting bit
CB2.4 to 1
bin
.
1
bin
Register lock active:
• The KL3356 no longer updates the registers
• Bit CB2.4 is not evaluated!
normal operation
manual operation*
CB2.2 MapCaliData/
DisableSymm
0
bin
The measured value registers contain the mapped raw
data from the converters:
•
: the calculated weight
•
: the strain gauge measuring signal
•
: the strain gauge reference signal
•
: the last actual test value
Symmetrical
measurement is
switched on if it has
been enabled by bit
of the
feature register.
1
bin
The measurement registers contain the mapped
calibration data:
•
: reserved (register is empty)
•
: the offset error of the measuring signal
•
: the offset error of the reference signal
•
the correction factor for the differential
signal
Symmetrical
measurement is
switched off even if it
has been enabled by
bit
the feature register.
CB2.1 CaliDisReq/
Channel
0
bin
Selection of the
measuring channel,
U
Diff
1
bin
Blocking automatic calibration
Selection of the
measuring channel,
U
Ref
CB2.0 StartManCheck/
StartManCali
1
bin
Start test
Start calibration and
test
*) Manual operation can be enabled via the command register (
Status byte 2 (for process data mode)
The status byte 2 (SB2) is located in the
and is transmitted from terminal to the controller.
Bit
SB2.7 SB2.6 SB2.5 SB2.4
SB2.3
SB2.2
SB2.1
SB2.0
Name
-
Error
-
ForcedCali RegLockAck NoActualValue
CaliDisAck/Channel NegWeight