6
PO8C User Guide
Pressure Regulator Specific Warnings and Notes
Air supply: If the drain removal from the air filter and mist
separator is missed, drain will be flown to the regulator (and
onwards), resulting in a possibly malfunctioning equipment.
When removing drain is difficult, use of a filter with auto-
drain is recommended.
If the supply pressure line contains drain or particlate etc.
the pressure regulator's fixed throttle can become clogged
leading to malfunction. Therefore, in addition to an air filter,
be sure to use a mist separator.
If you use nitrogen, minimize the leak to the atmosphere and
take care of sufficient ventilation. Please note the Pressure
Adjuster's Bleed Flow and Exhaust Flow rates found among
the Specifications starting from page 3. Close the valve of
the nitrogen cylinder, when the system is not in use.
Increase in the percentage of nitrogen in the ambient air may
cause unconsciousness and death without warning. Read
carefully the safety instructions for nitrogen and make sure
that the other people in the same space are aware of the
danger.
Do not use the regulator outside the range its specifications
as this can cause failure. Specifications can be found on
page 3 of this manual.
Air is normally released from the bleed hole (the hole in the side
of the body's mid-section). This is a necessary consumption of air
based on the construction of the precision regulator and is not an
abnormality.
When changing the set pressure, keep in mind that changes in
gas pressure affects the gas temperature (which again, affects
the pressure). Also, the tension/volume of the measurement
hoses vary depending on the set pressure.