BEAMEX PGC Скачать руководство пользователя страница 1

 

INSTRUCTION MANUAL 

 

BEAMEX

®

 PRESSURE/VACUUM  

PUMP MODEL PG

 

  

 

Содержание PGC

Страница 1: ...INSTRUCTION MANUAL BEAMEX PRESSURE VACUUM PUMP MODEL PGC ...

Страница 2: ......

Страница 3: ...ility for any errors in this manual or their eventual consequences We reserve rights to make modifications to this manual without any further notice For more detailed technical data about the Instruction manual for PGC Calibration Pressure Pump please contact the manufacturer 2016 BEAMEX OY AB Ristisuonraitti 10 FIN 68600 Pietarsaari FINLAND Tel 358 10 5505000 Fax 358 10 5505404 E mail sales beame...

Страница 4: ... Accessories 1 USING THE PUMP 3 Preparation 3 Operation Pressure 4 Operation Vacuum 5 When Calibration is Done 5 TROUBLESHOOTING MAINTENANCE 6 Seal Replacement 6 SPECIFICATIONS 8 The Pump Unit 8 The Pressure Measurement T hose Part of Pump Kit 8 WARNINGS 9 ...

Страница 5: ...MC6 Calibrator s internal high pressure modules G 1 8 male 60 internal cone G 1 4 B female 2 seals An adapter for connecting the pressure measurement T hose to Beamex high pressure EXT modules G 1 8 male 60 internal cone G 1 4 NPT male An adapter for connecting the pressure measurement T hose to the instrument to be calibrated DUT 40 bar measurement T hose Carrying case Optional Accessories The op...

Страница 6: ...2 Presenting PGC Device Under Test DUT Reference meas urement Connection diagram when calibrating ...

Страница 7: ...hose to connect the Device Under Test DUT and the reference a calibrator with internal pressure modules or an external pressure module communicating with a calibrator to the pump For help on connections refer to the adjacent connection diagram Use hand tightening for the connections Ensure that the measurement range of the connected reference is appropriate Note It is not recommended to use the pl...

Страница 8: ...est pressure value is reached and there is no need for calibration points with decreasing pressure either Open the release valve close it and start a new calibration repeat or continue to chapter When Calibration is Done on page 5 If there is need for calibration points with decreasing pressure carefully open the release valve to decrease the pressure When getting close to the required pressure cl...

Страница 9: ... new calibration repeat or continue to chapter When Calibration is Done on page 5 If there is need for calibration points with increasing pressure carefully open the re lease valve to increase the pressure When getting close to the required pressure close the release valve and use the fine adjustment valve to increase the pressure to the calibration point Again wait for approximately 30 seconds fo...

Страница 10: ...and vacuum side If so push it in place as needed If a pressure increase decrease still cannot be obtained it s possible that one or more of the seals in the Pressure Pump is leaking and needs to be replaced If the test pump has not been used for a longer period of time the first lift may be somewhat sluggish This effect will disappear again during further operation Do not continue pumping if the f...

Страница 11: ...7 ...

Страница 12: ...Pressure range 0 95 to 35 bar 13 7 to 510 psi Pressure media Air Output connector G 1 8 male 60 internal cone Material Aluminium brass ABS NBR The Pressure Measurement T hose Part of Pump Kit Max length 1 5 m 4 9 ft Connectors all ends G 1 8 female 60 internal cone External diameter 5 mm 0 2 Operating pressure max 40 bar 4 MPa 580 psi ...

Страница 13: ...he pressure vacuum selector s setting when there is pressure or vacuum in the pump system Vent external systems before connecting to the pump Ensure that all connections are made correctly and that the hose and the con nectors are undamaged Do not use faulty hoses or connectors The environmental conditions may restrict the allowable maximum pres sure vacuum to a lower level than the pump and the h...

Страница 14: ...10 ...

Страница 15: ......

Страница 16: ...30067 U S A Phone 800 888 9892 1 770 951 1927 Fax 1 770 951 1928 E mail beamex inc beamex com Beamex Limited Newtown Grange Farm Business Park Desford Road NEWTOWN UNTHANK Leicestershire LE9 9FL United Kingdom Phone 01455 821 920 Fax 01455 821 923 E mail beamex ltd beamex com Beamex S A S 253 Boulevard de Leeds 59777 Lille FRANCE Phone 33 0 3 28 53 58 27 Fax 33 0 3 28 53 57 50 E mail beamex fr bea...

Отзывы: