Measurements - 45
Voltage Measurement
MC6-T's voltage measurement terminals are listed below as they are
shown in
Figure 28: Voltage measurement terminals
•
TC1
, measurement range: -1.01 to +1.01 VDC.
•
TC2
, measurement range: -1.01 to +1.01 VDC.
•
IN
, measurement range: -1.01 to +60.6 VDC.
Figure 28: Voltage measurement terminals
Note that you may measure non-supported thermocouple signals using
either
TC1
or
TC2
port. The reading is in (milli)volts, so you need a data
table to convert the signal to temperatures.
Scaling
tool may be used for
converting millivolts to temperatures.
Warning:
Do not apply hazardous voltages to MC6-T's
terminals.
See also chapters
Frequency Measurement
In frequency measurement, be sure you select a suitable
trigger level
setting. To do that, tap on the button
. From the opened pop-up
window: Select a suitable trigger level.