66
Beam®, Operating Manual, Rev. 4.6.1 01/2016, GmbH
Error tables
6
6.5 Error tables of ECG-Recorders Beam
®
Error Symptoms
Possible Causes
Remedy
Continuous red flashing of
the LED every 2 minutes
and a three-tone audible
signal
Error message Priority 1
Fatal Error
If the error still appears after
replacing the battery, please
send the device to your ven-
dor or directly to IEM GmbH
for testing and repair.
every 2 seconds red
flashing of the LED, every
2 minutes a 3-tone audible
signal.
Error message Priority 2
Main battery discharged
Replace battery.
every 5 seconds a 2-tone
red flashing of the LED,
every 5 minutes a 3-tone
audible signal.
Error message Priority 3
Single-use electrodes in
the Loop mode are not
connected
■
Check that the cable
plug is securely attached
to the device port.
■
Check that the adhesive
electrodes are securely
attached to the patients
chest.
■
Only use single-use
electrodes of branded
manufacturers.
■
Plug the single-use elec-
trodes first onto the
cable and then posi-
tion the electrodes on
the patients chest. The
contact gel must not be
squeezed out as result
of too much pressure.
every 6 seconds a red 3-
tone flashing of the LED,
every 2 minutes a 3-tone
audible signal.
Error message Priority 3
Memory is full
Assign the values to the
respective patient in the HMS
CS and delete the memory of
the ECG recorder Beam
®.