background image

Part No. 847706-00 Rev. C00 Pg. 12

©2012 BeaconMedæs.  All rights reserved.

Replacement Components

DISS Outlets

1

2

1

2

DISS Concealed Wall Outlet

DISS Console Outlet

Item

Part Number

Description

1.

---

Piston (Not available as separate item.  Order appropriate repair kit below.)

2.

O-ring

622585-PG

For Oxygen, Vacuum, Evacuation and WAGD (pkg of 10)

622586-PG

For N2O, Air, N2 and CO2 (pkg of 10)

Refer to page 6 for complete replacement assemblies.

Repair Kits:  

(To repair 10 outlets, order 1 repair kit for appropriate gas service.)

Part Number

Description

290805-PG

Primary Check Repair Kit (Oxygen)

290806-PG

Primary Check Repair Kit (N2O, Air, N2 and CO2)

290807-PG

Primary Check Repair Kit (Vacuum and Evacuation)

290808-PG

Primary Check Repair Kit (WAGD)

NOTE:

 

Repair Kits are not interchangeable 
between EVACUATION and WAGD 
services.  Be certain to order correct 
repair kit.

Содержание Diamond II

Страница 1: ...Service Manual Gemini Diamond II and Diamond III Medical Gas Outlets for Concealed Piping and Console Applications MAN01 049 Part number 847706 00 Rev C00...

Страница 2: ...eded to service BeaconMed s Gemini Diamond II and Diamond III Gas Outlets Information contained in this manual is intended for use by technicians or personnel quali ed to repair and service medical eq...

Страница 3: ...vice WARNING Do not use oil on adapters or outlets Gemini Concealed Quick Connect Front Body DISS outlets are typically used in ceilings of operating rooms or in place of quick connect outlets Connect...

Страница 4: ...s compatible with acetal and nylon plastic anodized aluminum and die cast zinc Lubricate elastomer seals sparingly with Krytox GPL206 Refer to Tools and Service Materials DO NOT USE OIL Inspection and...

Страница 5: ...le outlets will appear as shown above Front body will have a solid rectangle of color coded plastic approximately 1 7 8 wide x 2 3 4 tall Raised circular keying portion of front body will rise approxi...

Страница 6: ...2 00 234712 00 Vacuum 234513 00 234713 00 Nitrogen 234514 00 234714 00 WAGD 234519 00 234719 00 Carbon Dioxide 234520 00 234720 00 Gemini Console Quick Connect Oxygen 234410 00 234810 00 Nitrous Oxide...

Страница 7: ...ement is illustrated in exploded parts breakdown pages 10 12 NOTE Gas supply need not be turned off when removing front body secondary check and bushing in back body prevents ow from outlet Adapter lo...

Страница 8: ...ace Faulty DISS nipple Faulty O Ring seal Replace DISS nipple Replace O Ring seal Adapter will not engage and or release Damaged adapter Replace adapter Wrong adapter Check adapter outlet gas indexing...

Страница 9: ...5344 00 Secondary Check each 3 4107 2097 38 O Ring Black pkg of 10 4 622601 PG Rubber Bushing pkg of 10 5 4107 2097 88 Retaining Ring pkg of 10 Repair Kits Below Repair Kits Contain Components to Repa...

Страница 10: ...6 PG Spring pkg of 10 2 4107 2097 44 Piston pkg of 10 3 622589 PG O Ring Red pkg of 10 4 415038 PG Washer pkg of 10 5 833953 PG Lock in Adapter Spring pkg of 10 6 4107 2097 31 Eyelet pkg of 50 7 4107...

Страница 11: ...3436 PG Spring pkg of 10 2 4107 2097 44 Piston pkg of 10 3 622589 PG O Ring Red pkg of 10 4 415038 PG Washer pkg of 10 5 833953 PG Lock in Adapter Spring pkg of 10 6 4107 2097 31 Eyelet pkg of 50 7 41...

Страница 12: ...pkg of 10 622586 PG For N2O Air N2 and CO2 pkg of 10 Refer to page 6 for complete replacement assemblies Repair Kits To repair 10 outlets order 1 repair kit for appropriate gas service Part Number De...