6-7
“Oil-Less” Scroll Laboratory Air
4107 9006 25.11
Figure 6.6 Kit_001
6.6 Backup Pressure Switch
Set Point Adjustments
The backup switch is set at the factory to the
operating point(s) as stated on the wiring
diagram supplied with the unit. It is good
practice to cycle the switch to determine
actual operating points before proceeding
with readjustment.
POS
Description
Qty
01
Tank and level sensor
-
02
Electronic board
-
03
Solenoid valve
-
04
Filter cap
1
05
Filter cap O-Ring
1
06
Double mesh metallic
fi
lter
1
07
Valve assembly O-Ring
1
08
Solenoid valve spring
1
09
Solenoid valve pilot
1
10
Diaphragm
1
11
Valve assembly
1
12
M6x25 screws
4
13
M3x10 screws
4
14
Flow limiter
1
15
Flow limiter plastic ring
1
CAUTION:
• ALWAYS change pressure setting
gradually.
• ALWAYS check switch setting before
making any adjustments.
• DO NOT force slotted adjustment screw
when it becomes di
ffi
cult to turn.
• ALWAYS isolate the pressure transducer
before making any adjustments to the
backup pressure switch.
Exploded view: