Signal detectors
Light signals are generated as particles pass through the laser beam in a fluid stream. When these optical
signals reach a detector, electrical pulses are created that are then processed by the electronics system.
There are two types of signal detectors in the BD FACSymphony™ A1 flow cytometer:
l
Photomultiplier tubes (PMTs).
Used to detect the weaker signals generated by side scatter and all
fluorescence channels. These signals are amplified by applying a voltage to the PMTs.
l
Photodiodes.
Less sensitive to light signals than the PMTs. A photodiode is used to detect the stronger
forward scatter (FSC) signal.
More information
l
Optical filter theory (page 80)
l
About configurations (page 100)
Workstation
Introduction
This topic describes the components of the BD FACSymphony™ A1 workstation.
Workstation components
Acquisition, analysis, and most instrument functions are controlled by the BD FACSymphony™ A1 workstation.
It includes a PC and one or two monitors.
Your workstation is equipped with the following:
l
Microsoft Windows operating system
l
BD FACSDiva™ software version 9.0.2 or later for data acquisition and analysis
l
Software documentation including the help system
More information
l
About the BD FACSymphony™ A1 documentation (page 8)
Chapter 2 Introduction
23
Содержание FACSymphony A1 Flow
Страница 1: ...FACSymphony A1 Flow Cytometer User s Guide 23 23437 01 2022 07 For Research Use Only ...
Страница 6: ......
Страница 10: ...This page intentionally left blank ...
Страница 24: ...This page intentionally left blank ...
Страница 36: ...This page intentionally left blank ...
Страница 50: ...This page intentionally left blank ...
Страница 87: ...More information l Running a performance check page 55 Chapter 7 Technical overview 87 ...
Страница 88: ...This page intentionally left blank ...
Страница 96: ...This page intentionally left blank ...
Страница 120: ...This page intentionally left blank ...