Droplet containment module
The DCM prevents sheath fluid from dripping from the SIP and provides biohazard protection.
The DCM vacuum is activated when the tube support arm is moved to the side. Sheath fluid is aspirated as it
backflushes the sample injection tube. This backflush helps prevent carryover of cells between samples.
Sheath and waste tanks
Introduction
This topic describes the sheath and waste tanks. The sheath and waste tanks are outside the cytometer and
can be positioned on the floor.
Note:
Only BD service engineers should change the location of the sheath and waste tanks.
If you are using the BD FACSFlow™ supply system (FFSS), see the BD FACSFlow™ Supply System User’s Guide
for more information.
Sheath tank
The sheath tank has a capacity of 10 L. Sheath fluid is filtered through an in-line, interchangeable filter that
prevents small particles from entering the sheath fluid lines. An alarm sounds when the tank is empty.
Do not fill the sheath tank to its maximum capacity. When an overfull tank is pressurized, erratic
cytometer performance can result.
Waste tank
The waste tank has a capacity of 10 L. An alarm sounds when the tank is full.
Cable management
Arrange the cables, air lines, and fluidics lines from the instrument to the sheath and waste tanks away from
high traffic area. Also ensure that the cables and lines are placed so they do not interfere with normal
instrument operation.
Routing the cables and lines from the back of the instrument to the underside of the table and then to the
tanks is preferable. Any excess lines or tubing should be loosely coiled but out of the way of the tanks.
More information
l
Preparing the sheath tank (page 27)
l
Preparing the waste tank (page 30)
l
l
Fluidic alarms and the Mode button (page 19)
Chapter 2 Introduction
21
Содержание FACSymphony A1 Flow
Страница 1: ...FACSymphony A1 Flow Cytometer User s Guide 23 23437 01 2022 07 For Research Use Only ...
Страница 6: ......
Страница 10: ...This page intentionally left blank ...
Страница 24: ...This page intentionally left blank ...
Страница 36: ...This page intentionally left blank ...
Страница 50: ...This page intentionally left blank ...
Страница 87: ...More information l Running a performance check page 55 Chapter 7 Technical overview 87 ...
Страница 88: ...This page intentionally left blank ...
Страница 96: ...This page intentionally left blank ...
Страница 120: ...This page intentionally left blank ...