Setup Mode displays the current settings of Sound, FFSS Mode, and Altitude while the system is running. The
Setup Mode screen is shown in the following illustation.
SETUP
ALTITUDE
(METERS)
SOUND
FFSS
SW Revision Ø6
2000
1500
1000
500
0
-500
OFF
EXIT SETUP
4
3
2
1
Number
Definition
1
Sound.
Turns the warning alarm sounds on or off. To turn off the alarm, slide the slider to the left position.
To turn on the alarm, slide the slider to the right position. A green light indicates the alarm is active. A red
light indicates the alarm is inactive.
2
FFSS Mode.
Turns FFSS Mode on or off. To turn off FFSS Mode, slide the slider to the left position. To turn on
FFSS Mode, slide the slider to the right position. A green light indicates FFSS Mode is active. A red light
indicates FFSS Mode is inactive.
3
Exit Setup.
Button to exit Setup Mode.
4
Altitude.
Shows the current setting of Altitude. The slider enables quick changes to the Altitude setting by
moving the slider up or down. Altitude can be set from -500 to 2000 meters. Sliding the slider down to the
Off position turns the air pressure warning off.
Fluidic alarms and the Mode button
The fluidic alarms are triggered by the waste and sheath fluid levels in the tanks and by low or high sheath tank
air pressure levels. The alarms sound when the waste tank is nearly 100% full and the sheath tank is empty. The
fluidic alarms and system status will also show warnings when you start up the cytometer until the tank
pressure reaches the correct level.
To silence the alarm sounds, press the
Setup Mode
in the touchscreen. Slide the
Sound
slider to the left. To turn
the alarm sounds back on, slide the
Sound
slider to the right. If the alarm sounds are turned off, the
Normal
Mode
screen will also indicate the sounds are off.
Note:
The system status indicators are not affected by the alarm sounds on/off state.
Turning off the sound may result in overflow of waste and may result in running out of sheath fluid.
Note:
When the cytometer is in FFSS mode, both visual and audible alarms are deactivated.
Note:
The function of the system is sensitive to air pressure for flow rates. The system needs to know the
altitude, in meters, of the location of the system in order to accurately report the air pressure.
Chapter 2 Introduction
19
Содержание FACSymphony A1 Flow
Страница 1: ...FACSymphony A1 Flow Cytometer User s Guide 23 23437 01 2022 07 For Research Use Only ...
Страница 6: ......
Страница 10: ...This page intentionally left blank ...
Страница 24: ...This page intentionally left blank ...
Страница 36: ...This page intentionally left blank ...
Страница 50: ...This page intentionally left blank ...
Страница 87: ...More information l Running a performance check page 55 Chapter 7 Technical overview 87 ...
Страница 88: ...This page intentionally left blank ...
Страница 96: ...This page intentionally left blank ...
Страница 120: ...This page intentionally left blank ...