The RUN button will also turn orange if the sample tube is cracked or the BAL seal is bad. See
.
l
STANDBY
Stops fluid flow to conserve sheath fluid.
When you leave the cytometer for more than a few minutes, place a tube containing less than 1 mL of
deionized (DI) water on the sample injection port (SIP) and press STANDBY.
l
PRIME
Prepares the fluidics system by draining and filling the flow cell with sheath fluid.
Note:
Remove sample tubes during the prime cycle.
The fluid flow initially stops, and pressure is reversed to force fluid out of the flow cell and into the waste
container. After a preset time, the flow cell fills with sheath fluid at a controlled rate to prevent bubble
formation or entrapment. At completion, the cytometer switches to standby mode.
Sample flow rate control
The three flow rate buttons (LOW, MED, HIGH) control the sample rate through the flow cell. The three fluidic
control buttons (RUN, STANDBY, PRIME) enable you to control the sample acquisition rate.
2
1
Number
Component
1
Sample flow rate buttons
2
Sample fine adjust buttons
When sample adjust is set to 250 (as shown on the status screen on the control panel) the sample flow rates at
the Low, Med, and High settings are approximately 12, 35, and 60 µL/min of sample, respectively. Each time
you press one of the SAMPLE ADJ buttons, the fine adjust of the indicated sample increases or decreases by 10.
The following table shows the approximate sample flow rate range for low, medium, and high.
Settings
Sample flow rate (µL/min)
Low
6–24
Med
17.5–70
High
30–120
Chapter 2 Introduction
17
Содержание FACSymphony A1 Flow
Страница 1: ...FACSymphony A1 Flow Cytometer User s Guide 23 23437 01 2022 07 For Research Use Only ...
Страница 6: ......
Страница 10: ...This page intentionally left blank ...
Страница 24: ...This page intentionally left blank ...
Страница 36: ...This page intentionally left blank ...
Страница 50: ...This page intentionally left blank ...
Страница 87: ...More information l Running a performance check page 55 Chapter 7 Technical overview 87 ...
Страница 88: ...This page intentionally left blank ...
Страница 96: ...This page intentionally left blank ...
Страница 120: ...This page intentionally left blank ...