Chapter 4: Maintenance
37
14. Reconnect the air and fluid lines coming from the sheath
plenum.
More information
•
Flushing the BD LSRFortessa system (page 30)
Replacing the air filter
Introduction
This topic describes the procedure for replacing the air filters. An
air filter is located on the cap assemblies of the sheath and waste
cubitainers.
Procedure
To replace an air filter:
1. Remove the air filter.
Hold the silicone tubing with one hand and pull off the filter
with the other hand. See the following figure.
Caution: Biohazard!
Treat a contaminated air filter as
biohazardous waste.