Excitation Optics
The excitation optics consist of lasers, fiber optic cables, beam-shaping prisms,
and an achromatic focusing lens, as shown in Figure 1-3.
The BD FACSCanto II instrument uses low-powered air-cooled and solid state
lasers that do not have special power and cooling requirements.
Fiber optic cables direct the laser light onto beam-shaping prisms, which in turn
transmit the laser light to a focusing lens. The lens directs the laser light onto the
sample stream within the flow cell (Figure 1-3). See also Figure A-3 on page 109.
Figure 1-3
Optical pathway (figure shows violet laser available with 4-2-2 configuration only)
When the flow cell access door opens, an interlock shutters the laser light and
blocks its pathway for safety reasons.
Table 1-1
4-2 configuration fluorochromes
Laser
Wavelength
(nm)
Commonly Used Fluorochromes
Coherent® Sapphire™
Solid State
488 (blue)
FITC, PE, PE-Cy7,
PerCP or PerCP-Cy5.5
JDS Uniphase™ HeNe
Air Cooled
633 (red)
APC, APC-Cy7
1
2
3
4
interlock
focusing lens
prisms
flow cell
1
3
2
4
Содержание FACSCanto II
Страница 4: ......
Страница 12: ...xii BD FACSCanto II Flow Cytometer Reference Manual...
Страница 44: ...44 BD FACSCanto II Flow Cytometer Reference Manual...
Страница 104: ...104 BD FACSCanto II Flow Cytometer Reference Manual...
Страница 124: ...124 BD FACSCanto II Flow Cytometer Reference Manual...
Страница 138: ...138 BD FACSCanto II Flow Cytometer Reference Manual...
Страница 144: ...144 BD FACSCanto II Flow Cytometer Reference Manual...
Страница 152: ...152 BD FACSCanto II Flow Cytometer Reference Manual...