Table 3: Troubleshooting

BD EleVation™ 

Breast Biopsy System

Condition

Action Required

ERROR indicator blinks red, PRIME/PIERCE and 
SAMPLE buttons are solid red

Error during:

- Calibration

- PRIME/
PIERCE

- Sampling

- Remove the BD EleVation™ Probe

- Place BD EleVation™ Driver on a 

flat surface with hand away from 
any moving components

- Press SAMPLE button

ERROR indicator blinks red, SMART MODE indicator 
blinks red, PRIME/PIERCE and SAMPLE buttons 
are solid red

SMART MODE 
overloaded

- Remove BD EleVation™ Probe

- Place BD EleVation™ Driver on a 

flat surface with hand away from 
any moving components 

- Press SAMPLE button 

ERROR indicator blinks red, BATTERY indicator 
blinks red, PRIME/PIERCE and SAMPLE buttons 
are solid red

BD EleVation™ 
Driver Placed 
in wireless 
charging 
stand while 
BD EleVation™ 
Probe is loaded

- Remove BD EleVation™ Driver 

from wireless charging stand

- Remove BD EleVation™ Probe 

Table 4: Specifications

Environmental 
Conditions – Storage

Transport and store in a cool dry place.
Temperature:  -18ºC to + 40ºC
Humidity:  10% - 90% 
Atmospheric Pressure:  500 hPa - 1060 hPa  

Environmental 
Conditions – Operation

Operate the device in ambient conditions.
Temperature:  15ºC to 25ºC
Humidity:  30% - 75%
Atmospheric Pressure:  700 hPa - 1060 hPa

Ingress of Water

Not protected against the ingress of water. IPX0.

Flammability

Equipment not suitable for use in the presence of flammable anesthetic.

Dimensions

Approximately 4 cm x 6 cm x 25 cm 

Weight

Approximately 0.4 kg

AC Power Accessories 
and Ratings

Use only with supplied AC power accessories.
Power Input:
Input Voltage Range: 100~240 Vac
Input Frequency Range: 50~60Hz
Output Voltage: +5Vdc
Output Current Max: 2.4A

Removing the AC adapter from wall power shall serve as isolation mains.  Do not 
position the AC adapter and charging stand such that it is difficult to remove the AC 
adapter from the wall outlet if needed to remove mains power. 

Battery Type

Lithium Ion

Applied Part

BD EleVation™ Probe

Electrical Conformity

This medical equipment has passed all required testing for electric shock, fire and 
mechanical hazards in accordance with IEC/EN 60601-1, and CAN/CSA C22.2 No 
601-1. 

Classification

Class II, internally powered Type BF equipment

Mode of Operation

Continuous operation

Electromechanical 
Compatibility

Medical Electrical Equipment needs special precautions regarding EMC and needs to 
be installed and put into service according to the EMC information provided below. 

Portable and Mobile RF Communications Equipment can affect Medical Electrical 
Equipment.

Table 5: Guidance and Manufacturer’s Declaration – Emissions

The BD EleVation™ Breast Biopsy System is intended for use in the electromagnetic environment specified below. 
The user of the BD EleVation™ Breast Biopsy System should ensure that it is used in such an environment

Emissions Test

Compliance

Electromagnetic Environment – Guidance

RF Emissions
CISPR11

Group 1, Class A

The BD EleVation™ Breast Biopsy System 
uses RF energy only for its internal function. 
Therefore, its RF emissions are very low and 
are not likely to cause any interference in 
nearby electronic equipment.

Harmonic  
IEC 61000-3-2

Class A – Complies

The The BD EleVation™ Breast Biopsy 
System is suitable for use in all 
establishments, 

other than

 domestic, and 

those directly connected to the public low-
voltage power supply network that supplies 
buildings used for domestic purposes.

Flicker 
IEC 61000-3-3

Complies

Содержание EleVation

Страница 1: ...BD EleVation Sistema de Biopsia Mam ria Instru es de utiliza o OH9DWLRQ Dansk BD EleVation Brystbiopsisystem Brugsanvisning Svenska BD EleVation br stbiopsisystem Bruksanvisning Suomi BD EleVation Ri...

Страница 2: ......

Страница 3: ...ation Probes Green EVSC Additional Sample Containers For use with all BD EleVation Probes N A EVCover Instrument Cover Indications for Use The BD EleVation Breast Biopsy System is indicated to obtain...

Страница 4: ...psy System should only be used by a physician trained in its indicated use limitations and possible complications of percutaneous needle techniques 2 Do not attempt to remove the cover or modify the d...

Страница 5: ...sterile barrier or foreign material NOTE The BD EleVation Driver automatically recognizes when a BD EleVation Probe has been properly inserted and the BD EleVation Driver will perform a calibration t...

Страница 6: ...cluded with the BD EleVation Probe The BD EleVation Half Notch Support Cannula will reduce the length of the sample notch from 2cm to 1cm 5 To acquire multiple samples repeat Step 4 above NOTE Collect...

Страница 7: ...the wireless charging stand will illuminate green 4 Place the BD EleVation Driver in the wireless charging stand as shown in Figure 14 The BATTERY indicator on the BD EleVation Driver will illuminate...

Страница 8: ...Hz Output Voltage 5Vdc Output Current Max 2 4A Removing the AC adapter from wall power shall serve as isolation mains Do not position the AC adapter and charging stand such that it is difficult to rem...

Страница 9: ...5 V1 P D 3 5 E1 P 80 to 800 MHz D 7 E1 P 800 MHz to 2 7GHz Where P is the max power in watts and D is the recommended separation distance in meters Field strengths from xed transmitters as determined...

Страница 10: ...HANDLING OR USE OF THIS PRODUCT Some states countries do not allow an exclusion of implied warranties incidental or consequential damages You may be entitled to additional remedies under the laws of y...

Страница 11: ...ation avec les sondes 12G BD EleVation Bleu EVH14 Canule de support demi encoche 14G Pour utilisation avec les sondes 14G BD EleVation Vert EV10S Stylet 10G Pour utilisation avec les sondes 10G BD Ele...

Страница 12: ...Vation Apr s rest rilisation la st rilit de la sonde BD EleVation n est pas garantie du fait d un degr ind termin de contamination pyrog ne ou microbienne potentielle pouvant provoquer des complicatio...

Страница 13: ...rer la sonde BD EleVation L indicateur BATTERY s allume en vert lorsque la charge est suffisante pour effectuer une biopsie Laisser charger le guide BD EleVation pendant la nuit apr s l avoir utilis...

Страница 14: ...u des l sions 4 Pr lever l chantillon en appuyant sur le bouton SAMPLE Le d roulement du processus de pr l vement est automatique a Le guide BD EleVation cr e un vide b La canule de coupe de la sonde...

Страница 15: ...eVation bien s cher l air libre avant de le placer dans la station de chargement sans l Hormis un nettoyage r gulier aucun entretien pr ventif n est requis Il est toutefois recommand de renvoyer le di...

Страница 16: ...pannage sp ci ques Si une condition d alerte persiste apr s avoir corrig des probl mes connus et remplac la sonde BD EleVation contactez un technicien agr NOTE si l erreur du syst me de biopsie mammai...

Страница 17: ...ions RF CISPR11 Groupe 1 Classe A Le syst me de biopsie mammaire BD EleVation utilise l nergie RF uniquement pour sa fonction interne Par cons quent ses missions RF sont tr s faibles et ne sont pas su...

Страница 18: ...3 6893 3 6893 7 3786 100 11 6666 11 6666 23 3333 Distances de s paration recommand es entre les quipements de communication RF portables et mobiles et le syst me biopsie mammaire BD EleVation quipeme...

Страница 19: ...Sonden Blau EVH14 14G Halbkerbenst tzkan le Zur Verwendung mit 14G BD EleVation Sonden Gr n EV10S 10G Mandrin Zur Verwendung mit 10G BD EleVation Sonden Wei nicht zutreffend EV12S 12G Mandrin Zur Verw...

Страница 20: ...kann nach einer erneuten Sterilisation nicht gew hrleistet werden da das Ausma einer evtl zu infekti sen Komplikationen f hrenden potenziellen Pyrogen oder Mikrobenkontamination nicht zu ermessen ist...

Отзывы: