Alaris™ Syringe Pumps
Controls and Indicators
Controls and Indicators
Controls:
Symbol
Description
a
ON/OFF
button - Press once to switch the Pump on . Press and hold down for 3 seconds to switch the
Pump off .
Note:
Logs are maintained for power down events including when the pump is powered down or
unexpected power loss .
b
RUN
button - Press to start the infusion . The green LED will flash during infusion .
h
HOLD
button - Press to put the infusion on hold . The amber LED will be lit while on hold .
R
MUTE
button - Press to silence alarm for two minutes . To re-enable the alarm audio press the
MUTE
button a second time .
Note:
Attention alarm only: when not in alarm press and hold until four audible beeps are heard for
15 minutes silence .
i
PURGE/BOLUS
button - Press to access
PURGE
or
BOLUS
soft keys . Press and hold down soft key to
operate .
PURGE
- primes the extension set with fluid or drug during initial set up .
• Pump is on hold
• Extension set must not be connected to the patient
• Volume Infused (VI) is not added
BOLUS
- fluid or drug delivered at an accelerated rate .
• Pump is infusing
• Extension set should be connected to the patient
• VI is added
d
OPTION
button - Press to access optional features, see 'Basic Features' section .
e
PRESSURE
button - Use this button to display the pumping pressure and alarm level .
This button will also display the pressure trend display .
f
CHEVRON
keys - Double or single for faster/slower increase or decrease of values shown on display .
g
BLANK SOFTKEYS
- Use in conjunction with the prompts shown on the display .
Indicators:
Symbol
Description
j
BATTERY
indicator - When illuminated the Pump is running on the internal battery . When flashing the
battery power is low with less than 30 minutes of use remaining .
S
AC POWER
indicator - When illuminated the Pump is connected to an AC power supply and the battery
is being charged .