BC Group International BC Biomedical ULT-2020 Series Скачать руководство пользователя страница 28

 

19 

System Configuration 

Parameter 

Description 

Options 

Test Mode 

Determines  whether  the  test  measurements  (actual 
test readings) are shown in the test screen, or if the 
unit  will  simply  give  a  PASS  or  FAIL  result.    The 
Default setting is Numerical. 

PASS / FAIL or  

Numerical 

Test Limits 

Determines the settings for the test limits.  Fixed limits 
are set by the factory and cannot be adjusted.  User 
limits are programmed by the user through a special 
access  code.    Manufacturer  Default  limits  are 
suggested  test  settings  by  probe  manufacturers.  
Custom limits are programmable by the user with the 
PC configuration program.     

Fixed, User, 

Manufacturer 

Default, or  

Custom 

DUT Manuf 

Selects the DUT manufacturer.   
NOTE:  Only available when Test Limit mode is set to 
Manufacturer Default or Custom.   

User Defined 

DUT Model 

Selects the DUT Model.   
NOTE:  Only available when Test Limit mode is set to 
Manufacturer Default or Custom. 

User Defined 

Allow User Config 

Selects whether the user is allowed to modify the DUT 
Manufacturer/Model  in  the  Device  Configuration 
Screen.  This allows an administrator to “Lock Down” 
the test configuration. 

yes/no 

Setup Clock 

Pressing  the  Up  arrow  while  this  parameter  is 
selected  displays  the  Clock  Configuration  screen 
where the Date and Time are configured.   

Press UP 

Printer 

Pressing  the  Up  arrow  while  this  parameter  is 
selected changes which model printer is selected. 

BC Group, Extech 

Erase Log 

Pressing  the  Up  arrow  while  this  parameter  is 
selected  will  erase  the  test  records  stored  in  the 
Datalog (ULT-2020 Only) 

Press UP 

Battery Life 

Displays the current battery life.   

0-100% 

(Read Only)

 

Contrast Adjust 

Sets the contrast of the display.   

0-20 

Backlight 

Off 

– Always off 

1-30 sec 

– The elapsed time after which the backlight 

will automatically turn off. 
ON 

– Always ON.   

The Default setting is 30 seconds. 

Off, 1-30 sec, ON 

Auto Off Timer 

Determines the period of inactivity before the meter is 
turned  OFF.  A  timer  is  started  when  the  meter  is 
turned  ON  and  is  reset  each  time  a  key  is  pressed. 
When  the  timer  reaches  the  value  set  in  this 
parameter, the power is automatically turned OFF.   
The Default setting is 15 minutes.   

 

(NOTE: Setting this parameter to 0 disables the Auto 
Off timer.) 

 

(NOTE:  Using a battery eliminator disables the Auto 
Off Timer.) 

0-30 minutes 

Access Code 

Sets  the  required  code  entry  to  gain  access  to  the 
Setup  Menu.    If  this  parameter  is  set  to  Zero,  the 
Access Code feature is disabled and the user will gain 
direct access to the Setup Menu. 

0-9999 

Software 

Displays current software program. 

(Read Only) 

Содержание BC Biomedical ULT-2020 Series

Страница 1: ...ULTRASOUND TRANSDUCER LEAKAGE TESTER ULT 2000 SERIES USER MANUAL...

Страница 2: ......

Страница 3: ...SCRIPTION 1 BACKGROUND 4 TEST PROCESS 8 LAYOUT 9 KEYS 10 SCREENS 11 SETUP 18 PC SOFTWARE 21 COMMUNICATION PROTOCOL 22 MANUAL REVISIONS 25 LIMITED WARRANTY 26 SPECIFICATIONS 27 NOTES 29 BC BIOMEDICAL U...

Страница 4: ...ed user modifications may result in hazards or improper operation CALIBRATION INTERVAL To ensure the accuracy of the ULT 2000 Series BC Group International Inc recommends calibration at least once eve...

Страница 5: ...l of 90 to 275 VAC WARNING CONNECTIONS All connections to patients must be removed before connecting the DUT to the ULT 2000 A serious hazard may occur if the patient is connected when testing with th...

Страница 6: ...sely affect the performance of the ULT 2000 Allow ULT 2000 to acclimate to specified conditions for at least 30 minutes before attempting to operate it CAUTION CLEANING Do not immerse The ULT 2000 sho...

Страница 7: ...v...

Страница 8: ...mputer Lbs pounds sec second s TEE Transesophageal Echocardiography ULT Ultrasound Transducer Leakage Tester USA United States of America V Volt s VAC Volts Alternating Current VDC Volts Direct Curren...

Страница 9: ...HIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE HOWEVER NO RESPONSIBILITY IS ASSUMED FOR INACCURACIES NOTICE CONTACT INFORMATION BC BIOMEDICAL BC GROUP INTERNATIONAL INC 3081 ELM...

Страница 10: ...RASOUND TRANSDUCER MANUFACTURER AND MODEL AUTO RANGING WITH 10 250 OR 500 A FS RANGES AUTOMATIC INTERNAL SELF TEST SINGLE BUTTON PRESS FOR FULL SYSTEM TEST SIMPLE PASS FAIL MODE FOR NON TECHNICAL USER...

Страница 11: ...M SOFT SIDED CARRYING Instrument and printer BC20 41361 COMMUNICATION CABLE USB TO MINI DIN M BC20 42200 CIDEX COMPATIBLE TEST BASIN BC20 42310 PACKAGE OF 5 ROLLS OF PAPER FOR PRINTER BC20 42322 CABLE...

Страница 12: ...ULT PA 27 GE Samsung GE 4C D 6VT D Voluson E8 Vivid E9 Samsung RS80A HS70A ULT PA 30 Hitachi HI VISION 900 5500 6500 8500 EUB 2000 EUB 525 EUB 405 Plus ULT PA 21 Philips ATL Mindray Philips iE33 and...

Страница 13: ...000 tests the integrity of the outer insulation barrier of the transducer and transducer cable as well as the existing capacitive leakage currents Due to the proximity of the TEE transducer to the hum...

Страница 14: ...can be the routine cleaning and disinfecting agent used as long as it is suitably conductive to electrical current flow Cidex is an example of a widely used disinfecting agent used for cleaning ultra...

Страница 15: ...h America the source voltage is typically set to 120 VAC 60 Hz For countries where the normal operating voltage is 230 VAC the source voltage can be set to this level at either 50 or 60 Hz as appropri...

Страница 16: ...G CONDUCTIVE SURFACES SUCH AS METAL CARTS OR TABLETOPS CAN CAUSE ERRONEOUS READINGS DUE TO ALTERNATE LEAKAGE PATHS FROM TRANSDUCER ADAPTERS CABLES ETC ENSURE TEST IS DONE ON A NON CONDUCTIVE SURFACE F...

Страница 17: ...y an alarm is activated and the test is halted Bath Conductivity Test The third step tests the conductivity of the liquid in the test basin The ULT 2000 accomplishes this task quite easily and reports...

Страница 18: ...ty and Leakage Scrolls Through Selected Parameter Options SELECT PRINT DATA Selects Next Available Parameter Prints Data MODE Changes Operating Mode Enters Exits Setup Mode BATH TEST DOWN Tests Conduc...

Страница 19: ...rints the latest test results to the serial port All other screens this key selects the next available parameter When not in a SETUP menu this key resets the system to the main screen When in a SETUP...

Страница 20: ...Manufacturer Model Progress Indicator Full Test Process SOURCE VOLTAGE TEST verifies the voltage circuit to be used during testing SELF TEST verifies that the ULT 2000 Series unit is working properly...

Страница 21: ...e shown below indicates that the probe leakage current is above the selected upper limit Bath Test Process SOURCE VOLTAGE TEST verifies the voltage circuit to be used during testing SELF TEST verifies...

Страница 22: ...uency and Output Control This screen is accessed from the Main Screen by pressing the key The Source Voltage Frequency and Output Control are selected by pressing the key until the desired parameter i...

Страница 23: ...surements and Test Limits This screen is accessed from the Meter Screen by pressing the key Scroll through the Datalog Records using the keys Test Results Pass Fail Test Limits 1 500 A Probe Informati...

Страница 24: ...15 Use the key to print the displayed record to the serial printer Below is a sample print...

Страница 25: ...d cannot be adjusted User Limits are user selectable Refer to USER TEST CONFIGURATION Manufacturer Default Limits suggested by the probe manufacturer and cannot be adjusted Custom Manufacturer Model a...

Страница 26: ...Eliminator BC20 21111 that may be used for continuous run applications It bypasses the internal battery when plugged in NOTE The unit is shipped with a Red Battery Lock Out plug installed into the lin...

Страница 27: ...ered the selected configuration See System Configuration and User Configuration for access code details When the desired access code is displayed press the key again to access the selected setup menu...

Страница 28: ...this parameter is selected displays the Clock Configuration screen where the Date and Time are configured Press UP Printer Pressing the Up arrow while this parameter is selected changes which model p...

Страница 29: ...of the Test Voltage when Device Configuration is set to User Test Limits 50 or 60 Hz Lkg Upper Limit The maximum allowable current leakage during a test when Device Configuration is set to User Test L...

Страница 30: ...Software then connect the PC and ULT 2000 via cable BC20 41361 See OPTIONAL ACCESSORIES at the beginning of this manual Once connected the software allows the user to download read custom configurati...

Страница 31: ...st position Further nodes are indented one position to the right of the parent node The highest level node of a command is called the Keyword followed by the Node s and then the value Keywords and Nod...

Страница 32: ...275 FREQuency 50 hz 60 hz OUTput OFF ON INITiate FULLtest BATHonly METermode ABORt SYSTem TIME hh mm 24hr mode only TFORmat AMpm 24hr DATE yy mm dd DFORmat MMddyy DDmmyy BATlife read only CONtrast 0...

Страница 33: ...64 Test Failed 7 128 Test Passed 8 256 9 512 10 1024 11 2048 Program Mode 12 4096 Meter Mode 13 8192 14 16384 15 32768 Calibration Mode FACTory VTOLerance 0 25 FVOLtage 90 275 VAC FCONductivity 0 500...

Страница 34: ...9 DT7331CF Euro Transformer Edited Rev 10 DT7331CG Specifications Updated Misc Edits Rev 11 DT7331CG Misc Edits Rev 12 DT7331CG Format Updated Printer cable part number updated Misc Edits Rev 13 DT733...

Страница 35: ...SSED OR IMPLIED INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE BC GROUP INTERNATIONAL INC IS NOT LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DA...

Страница 36: ...CONDUCTIVITY 0 5 500 A 1 FS CONNECTIONS PHYSICAL ENVIRONMENTAL DISPLAY 128 X 64 Pixels Graphical LCD White LED Backlight MEMORY SETUP EEPROM All parameters RETENTION 10 Years Retention w o Power CONST...

Страница 37: ...lent BATTERY ELIMINATOR 10 VDC 300mA BC20 21111 Universal BATTERY LIFE CONTINUOUS 100 Full Tests Note Backlight set to OFF OFF 1 year RS232 COMMUNICATIONS BAUD 115200 DATA BITS 8 START BITS 1 STOP BIT...

Страница 38: ...29 NOTES...

Страница 39: ...30 NOTES...

Страница 40: ...NAL INC 3081 ELM POINT INDUSTRIAL DRIVE ST CHARLES MO 63301 USA 1 800 242 8428 1 314 638 3800 www bcgroupintl com sales bcgroupintl com ULT 2000 Series User Manual 05 20 Rev 20 LM 7331X UM Copyright 2...

Отзывы: