Страница 3: ...SCRIPTION 1 BACKGROUND 4 TEST PROCESS 8 LAYOUT 9 KEYS 10 SCREENS 11 SETUP 18 PC SOFTWARE 21 COMMUNICATION PROTOCOL 22 MANUAL REVISIONS 25 LIMITED WARRANTY 26 SPECIFICATIONS 27 NOTES 29 BC BIOMEDICAL U...

Страница 4: ...ed user modifications may result in hazards or improper operation CALIBRATION INTERVAL To ensure the accuracy of the ULT 2000 Series BC Group International Inc recommends calibration at least once eve...

Страница 5: ...l of 90 to 275 VAC WARNING CONNECTIONS All connections to patients must be removed before connecting the DUT to the ULT 2000 A serious hazard may occur if the patient is connected when testing with th...

Страница 6: ...sely affect the performance of the ULT 2000 Allow ULT 2000 to acclimate to specified conditions for at least 30 minutes before attempting to operate it CAUTION CLEANING Do not immerse The ULT 2000 sho...

Страница 7: ...v...

Страница 8: ...mputer Lbs pounds sec second s TEE Transesophageal Echocardiography ULT Ultrasound Transducer Leakage Tester USA United States of America V Volt s VAC Volts Alternating Current VDC Volts Direct Curren...

Страница 9: ...HIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE HOWEVER NO RESPONSIBILITY IS ASSUMED FOR INACCURACIES NOTICE CONTACT INFORMATION BC BIOMEDICAL BC GROUP INTERNATIONAL INC 3081 ELM...

Страница 10: ...RASOUND TRANSDUCER MANUFACTURER AND MODEL AUTO RANGING WITH 10 250 OR 500 A FS RANGES AUTOMATIC INTERNAL SELF TEST SINGLE BUTTON PRESS FOR FULL SYSTEM TEST SIMPLE PASS FAIL MODE FOR NON TECHNICAL USER...

Страница 11: ...M SOFT SIDED CARRYING Instrument and printer BC20 41361 COMMUNICATION CABLE USB TO MINI DIN M BC20 42200 CIDEX COMPATIBLE TEST BASIN BC20 42310 PACKAGE OF 5 ROLLS OF PAPER FOR PRINTER BC20 42322 CABLE...

Страница 12: ...ULT PA 27 GE Samsung GE 4C D 6VT D Voluson E8 Vivid E9 Samsung RS80A HS70A ULT PA 30 Hitachi HI VISION 900 5500 6500 8500 EUB 2000 EUB 525 EUB 405 Plus ULT PA 21 Philips ATL Mindray Philips iE33 and...

Страница 13: ...000 tests the integrity of the outer insulation barrier of the transducer and transducer cable as well as the existing capacitive leakage currents Due to the proximity of the TEE transducer to the hum...

Страница 14: ...can be the routine cleaning and disinfecting agent used as long as it is suitably conductive to electrical current flow Cidex is an example of a widely used disinfecting agent used for cleaning ultra...

Страница 15: ...h America the source voltage is typically set to 120 VAC 60 Hz For countries where the normal operating voltage is 230 VAC the source voltage can be set to this level at either 50 or 60 Hz as appropri...

Страница 16: ...G CONDUCTIVE SURFACES SUCH AS METAL CARTS OR TABLETOPS CAN CAUSE ERRONEOUS READINGS DUE TO ALTERNATE LEAKAGE PATHS FROM TRANSDUCER ADAPTERS CABLES ETC ENSURE TEST IS DONE ON A NON CONDUCTIVE SURFACE F...

Страница 17: ...y an alarm is activated and the test is halted Bath Conductivity Test The third step tests the conductivity of the liquid in the test basin The ULT 2000 accomplishes this task quite easily and reports...

Страница 18: ...ty and Leakage Scrolls Through Selected Parameter Options SELECT PRINT DATA Selects Next Available Parameter Prints Data MODE Changes Operating Mode Enters Exits Setup Mode BATH TEST DOWN Tests Conduc...

Страница 19: ...rints the latest test results to the serial port All other screens this key selects the next available parameter When not in a SETUP menu this key resets the system to the main screen When in a SETUP...

Страница 20: ...Manufacturer Model Progress Indicator Full Test Process SOURCE VOLTAGE TEST verifies the voltage circuit to be used during testing SELF TEST verifies that the ULT 2000 Series unit is working properly...

Страница 21: ...e shown below indicates that the probe leakage current is above the selected upper limit Bath Test Process SOURCE VOLTAGE TEST verifies the voltage circuit to be used during testing SELF TEST verifies...

Страница 22: ...uency and Output Control This screen is accessed from the Main Screen by pressing the key The Source Voltage Frequency and Output Control are selected by pressing the key until the desired parameter i...

Страница 23: ...surements and Test Limits This screen is accessed from the Meter Screen by pressing the key Scroll through the Datalog Records using the keys Test Results Pass Fail Test Limits 1 500 A Probe Informati...

Страница 24: ...15 Use the key to print the displayed record to the serial printer Below is a sample print...

Страница 25: ...d cannot be adjusted User Limits are user selectable Refer to USER TEST CONFIGURATION Manufacturer Default Limits suggested by the probe manufacturer and cannot be adjusted Custom Manufacturer Model a...

Страница 26: ...Eliminator BC20 21111 that may be used for continuous run applications It bypasses the internal battery when plugged in NOTE The unit is shipped with a Red Battery Lock Out plug installed into the lin...

Страница 27: ...ered the selected configuration See System Configuration and User Configuration for access code details When the desired access code is displayed press the key again to access the selected setup menu...

Страница 28: ...this parameter is selected displays the Clock Configuration screen where the Date and Time are configured Press UP Printer Pressing the Up arrow while this parameter is selected changes which model p...

Страница 29: ...of the Test Voltage when Device Configuration is set to User Test Limits 50 or 60 Hz Lkg Upper Limit The maximum allowable current leakage during a test when Device Configuration is set to User Test L...

Страница 30: ...Software then connect the PC and ULT 2000 via cable BC20 41361 See OPTIONAL ACCESSORIES at the beginning of this manual Once connected the software allows the user to download read custom configurati...

Страница 31: ...st position Further nodes are indented one position to the right of the parent node The highest level node of a command is called the Keyword followed by the Node s and then the value Keywords and Nod...

Страница 32: ...275 FREQuency 50 hz 60 hz OUTput OFF ON INITiate FULLtest BATHonly METermode ABORt SYSTem TIME hh mm 24hr mode only TFORmat AMpm 24hr DATE yy mm dd DFORmat MMddyy DDmmyy BATlife read only CONtrast 0...

Страница 33: ...64 Test Failed 7 128 Test Passed 8 256 9 512 10 1024 11 2048 Program Mode 12 4096 Meter Mode 13 8192 14 16384 15 32768 Calibration Mode FACTory VTOLerance 0 25 FVOLtage 90 275 VAC FCONductivity 0 500...

Страница 34: ...9 DT7331CF Euro Transformer Edited Rev 10 DT7331CG Specifications Updated Misc Edits Rev 11 DT7331CG Misc Edits Rev 12 DT7331CG Format Updated Printer cable part number updated Misc Edits Rev 13 DT733...

Страница 35: ...SSED OR IMPLIED INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE BC GROUP INTERNATIONAL INC IS NOT LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DA...

Страница 36: ...CONDUCTIVITY 0 5 500 A 1 FS CONNECTIONS PHYSICAL ENVIRONMENTAL DISPLAY 128 X 64 Pixels Graphical LCD White LED Backlight MEMORY SETUP EEPROM All parameters RETENTION 10 Years Retention w o Power CONST...

Страница 37: ...lent BATTERY ELIMINATOR 10 VDC 300mA BC20 21111 Universal BATTERY LIFE CONTINUOUS 100 Full Tests Note Backlight set to OFF OFF 1 year RS232 COMMUNICATIONS BAUD 115200 DATA BITS 8 START BITS 1 STOP BIT...

Страница 38: ...29 NOTES...

Страница 39: ...30 NOTES...

Страница 40: ...NAL INC 3081 ELM POINT INDUSTRIAL DRIVE ST CHARLES MO 63301 USA 1 800 242 8428 1 314 638 3800 www bcgroupintl com sales bcgroupintl com ULT 2000 Series User Manual 05 20 Rev 20 LM 7331X UM Copyright 2...