background image

 

 

 

 
 

DIGITAL 

ULTRASOUND  

WATTMETER 

 
 
 
 
 
 
 
 
 
 
 

USP-100A

 

 

USER MANUAL 

Содержание USP-100A

Страница 1: ...DIGITAL ULTRASOUND WATTMETER USP 100A USER MANUAL...

Страница 2: ......

Страница 3: ...Notes 6 Line Battery Power 6 Battery Installation 6 Set Up 7 Calibration 7 Serial Computer Printer Interface 7 Programming Print Control 8 RS 232 9 Output Formats 10 Shipping Instructions for the USP...

Страница 4: ...IN ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT THE INFORMATION IN THIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE HOWEVER NO RESPONSIBILIT...

Страница 5: ...he water to reflect off the target and is then absorbed by the rubber lining The radiant power is directly proportional to the total downward force weight on the target This weight is then transferred...

Страница 6: ...pace around the instrument for ease of operation and keep away from radiant heat sources Never set any material on the USP 100A or place your hands or fingers on it while taking readings PANEL CONTROL...

Страница 7: ...nsducer surface for uniform wetting no air pocket or bubbles should be on its surface 8 Allow 5 minutes for the scale to stabilize With no ultrasonic power applied to the transducer press the On Zero...

Страница 8: ...it Place cone target assembly into test tank and in the small tube inside of the large tube Press the On Zero Off button to zero the unit 2 Place ultrasound transducer 1 8 inch into water above the ce...

Страница 9: ...the unit Place the weight on the arm of the cone target Within 3 seconds the unit should read 14 6 watts 2 watts or 1 0 grams 01 grams If this reading is significantly off the USP 100A needs to be re...

Страница 10: ...o determine if only stable non zero values will be printed Load Zero When set to Off the Auto print feature is disabled Content All of these features can be set On or Off Numeric data only Header Gros...

Страница 11: ...is operated from batteries the menu item is available and the default setting is Off To enable COM1 Power must be turned On Baud Rates of 60 1200 2400 4800 9600 and 19 200 are available Parity Parity...

Страница 12: ...yed weight SLZP Auto Print stable non zero weight and stable zero reading xP Interval Print x Print Interval 1 3600 seconds H Enter Print Header Lines Z Same as pressing Zero key T Same as pressing Ta...

Страница 13: ...clamp 4 Fasten the case lid onto the base after making certain there is nothing loose inside 5 The package used for shipping should be strong and large enough to allow for adequate packing material on...

Страница 14: ...er Transducer Operating Frequency 0 5 to 10 MHz Test Media Degassed Water Computer Interface RS 232 600 19200 Baud Default Baud Rate 2400 Power 12 VDC 500 mA 120 VAC Adapter Supplied Adapters for 240...

Страница 15: ...und propagation in water closely approximates that in tissues see UL 1 1981 AIUM NEMA Standard Publication The ultrasonic attenuation in water can be taken as a lower limit on the attenuation which wi...

Страница 16: ...limits shown in the following table for diagnostic ultrasound have been extracted from FDA Section 510 k guidance to manufacturers on submissions and clearance as of February 1993 Refer to the AIUM a...

Страница 17: ...manner as the density Therefore in ultrasound the two components of pressure one alternating and the other direct are present The average AC pressure per cycle is zero but the DC pressure of radiation...

Страница 18: ...less notified immediately If the instrument is damaged in any way a claim should be filed with the carrier immediately To obtain a quotation to repair shipment damage contact BC Group Final claim and...

Страница 19: ...17...

Страница 20: ...18...