Appendix A 

ULT-2000 Series Quick Start Guide (SonoSite Ultrasound) 

 10/2019 

Appendix A 

CS Medical TD 100 Details 

 

 

Using a TD 100 to high-level disinfectant your TEE probe, perform electrical leakage testing inside the TD 100: 

-

 

Press START on the TD 100 to initiate a disinfection cycle and insert your TEE probe.  

-

 

Insert the tip of the ULT-PC-31 Submersible Conductivity Probe until the white mark is at the reservoir. 

-

 

Enter probe number, operator number and lot number of disinfectant as prompted by the TD 100. 

-

 

Insert the bottle of disinfectant into the TD 100. Do not close the disinfectant door but wait for the bottle to drain into the TD 
100.  There is no need for a saline solution, as the disinfectant will work as the conductive solution. 

-

 

Once the bottle is fully emptied into the TD 100 you are ready to begin electrical leakage testing. 

 