CODE 

DISPLAY 

FUNCTION 

DESCRIPTION 

00 

NSR 

Normal Sinus 

NSR of 80 BPM 

01 

TACH 

Sinus Tachycardia 

NSR of 140 BPM 

02 

BRAD 

Sinus Bradycardia 

NSR of 30 BPM 

03 

ASYS 

Asystole 

Flat line signal 

04 A-FLTR 

Atrial 

Flutter 

Repeating sequence of 5 atrial beats and 1 ventrical 

beat for twelve seconds, followed by a repeating 

sequence of 3 atrial beats and 1 ventrical beat for six 

seconds, followed by a repeating sequence of 2 atrial 

beats and 1 ventrical beat for six seconds 

05 A-FIB 

Atrial 

Fibrillation 

Absence of P-wave, irregular P-R interval rate and a low 

level signal 

06 

2AVB1 

2nd Degree AV Block I 

80 BPM with increasing P-R interval for four beats  

(160, 220, 400, 470 ms) followed by a P wave without a 

QRS 

07 

2AVB2 

2nd Degree AV Block II 

80 BPM with Normal P-wave and P-R interval but 

wider QRS complexes

 

08 

3AVB 

3rd Degree AV Block 

80 BPM with P wave rate of 80 BPM and QRS rate of 

30 BPM 

09 PAC 

PAC 

Insert 

NSR of 80 BPM with Periodic Abnormal 25 % early P 

waves (Single Insert, return to  NSR) 

10 PVC1 

PVC1 

Insert 

NSR of 80 BPM with periodic left focus premature 

ventricular beats with 20% premature timing  

(Single Insert, return to  NSR) 

11 PVC2 

PVC2 

Insert 

NSR of 80 BPM with periodic right focus premature 
ventricular beats with 20% premature timing 

(Single Insert, return to  NSR) 

12 CPLT 

Couplet 

Insert 

NSR of 80 BPM with 2 Type 1 PVCs  

(Single Insert, return to  NSR) 

13 V-TAC  Ventricular 

Tachycardia 

160 BPM, No P-wave, 

Beats similar to Type 1 PVC 

14 V-FIB  Ventricular 

Fibrillation 

Irregular waveform with no real 

P-wave or clear R-R interval and a high signal level 

Содержание PSR-2200-DS

Страница 1: ...PROGRAMMABLE PATIENT SIMULATOR REMOTE CONTROL PSR 2200 DS USER MANUAL...

Страница 2: ......

Страница 3: ...1 DESCRIPTION 3 OVERVIEW 5 KEYS 6 SCREENS 7 MESSAGES 8 OPERATION 9 MANUAL REVISIONS 14 WARRANTY 14 SPECIFICATIONS 15 NOTES 16 BC BIOMEDICAL PSR 2200 DS TABLE OF CONTENTS...

Страница 4: ...E IN ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT THE INFORMATION IN THIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE HOWEVER NO RESPONSIBILI...

Страница 5: ...ATIONS AND ADJUSTMENTS LARGE GRAPHICS DISPLAY WITH BACKLIGHT DISPLAY BACKLIGHT WITH ADJUSTABLE TIMER PROGRAMMABLE AUTO SHUTDOWN TO CONSERVE BATTERY LIFE BC GROUP PSR 2200 DS PATIENT SIMULATOR REMOTE C...

Страница 6: ...f the PSR 2200 DS and gives descriptions of the elements that are present OVERVIEW 20 Light Touch Keys 25 Coiled Cable For connection to a PC or Patient Simulator LCD Graphical Display with Backlight...

Страница 7: ...5 List of Codes Battery Compartment 1 9V Battery Model and Serial Number Label...

Страница 8: ...returns parameters to their Standard settings These keys are used to select the values of adjustable parameters as accessed through the key and codes When not in the Special Function mode these keys c...

Страница 9: ...t Simulator The Backlight Time Screen allows for the setting of the Backlight Off Timer The Backlight can be Off set to 1 20 seconds or On The keys are used to choose the desired setting After 3 secon...

Страница 10: ...messages COMMUNICATION ERROR This message indicates that the simulator did not acknowledge a command that was sent to it If the cables are connected properly cycle power on the PSR 2200 DS and the Pat...

Страница 11: ...or on the simulator PRE DEFINED FUNCTION KEYS A fixed function can be loaded by pressing the desired key To make adjustments to the output selections and the Number Keys can be used to enter the code...

Страница 12: ...by a P wave without a QRS 07 2AVB2 2nd Degree AV Block II 80 BPM with Normal P wave and P R interval but wider QRS complexes 08 3AVB 3rd Degree AV Block 80 BPM with P wave rate of 80 BPM and QRS rate...

Страница 13: ...on the Patient Simulator The key returns the parameter to its Standard settings CODE DISPLAY FUNCTION DESCRIPTION ATRIAL 15 AT TACH Atrial Tachycardia NSR of 160 BPM 18 PED TACH Pediatric Tachycardia...

Страница 14: ...ST 0 8 0 7 0 6 0 5 0 4 0 3 0 2 0 1 05 0 0 05 0 1 0 2 0 3 0 4 0 5 0 6 0 7 0 8 mV CODE DISPLAY FUNCTION DESCRIPTION PACEMAKER 32 A PACE Atrial 80 BPM with Pacer Pulse at the start of each P wave 33 AV P...

Страница 15: ...n at 80 BPM CALIBRATION 78 ECG RATE Heart Rate Calibration 30 40 45 60 80 90 100 120 140 160 180 200 220 240 260 280 300 BPM 81 2 5 Hz Linearity Speed 2 5 Hz triangle waveform 82 ECG SIZE ECG Size Cal...

Страница 16: ...ATE OF SHIPMENT EXCLUSIONS THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR IMPLIED INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE...

Страница 17: ...cklit ENCLOSURE 5 98 x 3 27 x 1 28 Inches 152 x 83 x 32 5 mm ABS Plastic WEIGHT 1 Lbs 0 45 Kg FACE PLATE Lexan Back printed OPERATING RANGE 15 to 40 C STORAGE RANGE 20 to 65 C Electrical POWER Battery...

Страница 18: ...16 NOTES...

Отзывы: