background image

 

 

 

 

 
 

MULTI-PARAMETER  

PATIENT SIMULATOR 

 
 
 
 
 
 
 
 
 
 
 

PS-2200 SERIES

 

 

 

USER MANUAL 

Содержание PS-2200 series

Страница 1: ...MULTI PARAMETER PATIENT SIMULATOR PS 2200 SERIES USER MANUAL...

Страница 2: ......

Страница 3: ...PERFORMACE 33 BLOOD PRESSURE 37 SWAN GANZ 45 RESPIRATION 49 TEMPERATURE 51 TRAINING 53 SETUP 59 POWER UP SETTINGS 63 SpO2 OPTION 65 FETAL MATERNAL OPTION 67 CARDIAC OUTPUT OPTION 71 OUTPUT CONNECTION...

Страница 4: ...or DUT CONTACT INFORMATION BC BIOMEDICAL BC GROUP INTERNATIONAL INC PO BOX 25125 9415 GENTRY AVE ST LOUIS MO 63125 USA 1 800 242 8428 314 638 3800 www bcgroupintl com sales bcgroupintl com NOTICE BC G...

Страница 5: ...The PS 2200 Family makes viewing and selecting the desired waveforms and parameters quick and intuitive with all operational information being available at the same time on two cursor based graphic d...

Страница 6: ...MINI DIN CONNECTORS FOR BP CABLES 9 VOLT BATTERY POWER BATTERY LIFE DISPLAY LOW BATTERY INDICATOR AVAILABLE BATTERY ELIMINATOR DISPLAY BACKLIGHT FULL REMOTE OPERATION VIA RS 232 FLASH PROGRAMMABLE FOR...

Страница 7: ...T ARRHYTHMIAS FOUR GENERAL GROUPS SUPRAVENTRICULAR PREMATURE VENTRICULAR CONDUCTION MANUAL AND AUTOMATIC TRIGGERING PACEMAKER FUNCTIONS Seven different pacemaker waveforms may be simulated Additionall...

Страница 8: ...2 0 2 5 3 0 3 5 4 0 4 5 5 0 mV Lead II AUTOMATIC MODE R WAVE RATE 30 60 80 120 200 250 BPM R WAVE WIDTH 8 10 12 20 30 40 50 60 70 80 90 100 110 120 130 140 150 160 170 180 190 200 ms R WAVE AMPLITUDE...

Страница 9: ...anual or Automatic simulation of that sequence Each channel has independent settings STATIC 10 5 0 20 40 50 60 80 100 150 160 200 240 250 300 320 400 mmHg AUTOMATIC STATIC PRESSURE MODE 8 DYNAMIC WAVE...

Страница 10: ...mily of SpO2 finger simulators the system will provide a pulse synchronized SpO2 output for all NSR rates RATE 30 40 45 60 80 90 100 120 140 160 180 200 220 240 260 280 300 BPM SpO2 OUTPUT 80 90 97 FE...

Страница 11: ...uro Version OPTIONAL ACCESSORIES BC20 02300 FETAL MATERNAL OPTION BC20 02400 CARDIAC OUTPUT OPTION MSP 2100E FINGERSIM INTERFACE WITH BC20 21101 BATTERY ELIMINATOR EURO MSP 2100U FINGERSIM INTERFACE W...

Страница 12: ...PS 2200 10 This page intentionally left blank...

Страница 13: ...ry Compartment Back 7 Light Touch Keys for Selecting Parameters and Settings LEFT and RIGHT Curved Arrows for Moving through Parameters UP and DOWN Arrows for Scrolling through Options ENTER for Selec...

Страница 14: ...hlighting the parameter available for selection The keys change the options for the highlighted parameter The cursor begins flashing if the parameter has been changed The key selects the changed optio...

Страница 15: ...t line in the display to the first pacemaker waveform choice The key enters the machine performance testing category and changes the first line in the display to the first performance waveform choice...

Страница 16: ...the backlight will drain the battery if left on it will automatically shut off after a few seconds when running on battery power Note This time is selectable in the System Setup screen The intensity o...

Страница 17: ...ion Respiration waveforms are provided that have adjustable rates from 15 to 120 BrPM Breaths Per Minute as well as an Apnea 0 BrPM setting The signal is generated by a variation in the impedance in e...

Страница 18: ...n with independent outputs AHA and IEC color coded labels are located on the face of the unit to aid in connecting the corresponding U S and International Patient Leads High Level Output A high level...

Страница 19: ...this window and continue use of the unit If the battery is not replaced before the battery reaches a critical level 0 Battery Life the unit will shut down The percentage of life left in the batteries...

Страница 20: ...atic setting In each of these the unit will sequence through the full range of settings automatically at a fixed rate as selected in the Auto Step Time Parameter When in this mode the time remaining i...

Страница 21: ...essing the category key The display will resemble the following 30 BPM 40 BPM 45 BPM 60 BPM 80 BPM 90 BPM 100 BPM 120 BPM 140 BPM 160 BPM 180 BPM 200 BPM 220 BPM 240 BPM 260 BPM 280 BPM 300 BPM Auto A...

Страница 22: ...o scroll to the desired option Then is used to accept the new setting Auto Rate If the BPM parameter is set to AUTO the unit will automatically sequence through all of the BPM settings starting with 3...

Страница 23: ...n 3 5 or 12 lead configurations It has independent outputs for each signal lead referenced to the right leg There are 35 Arrhythmias available that model abnormal heartbeats The PS 2200 is placed into...

Страница 24: ...Premature Atrial PAC Auto Atrial PAC Man Nodal PNC Auto Nodal PNC Man PVC 1 Auto PVC 1 Man PVC 1 Early Auto PVC 1 Early Man PVC 1 R on T Auto PVC 1 R on T Man PVC 2 Auto PVC 2 Man PVC 2 Early Auto PVC...

Страница 25: ...ll to the desired option Then is used to accept the new setting NOTE While in the Arrhythmia Group choice screen the key may be used for a second time to jump directly to the arrhythmias choices for t...

Страница 26: ...ing PVC Type 1 9 NSR Continuous PVC 1 Man Standard Type 1 Premature Ventricular Contraction NSR of 80 BPM with periodic left focus premature ventricular beats with 20 premature timing One Time event P...

Страница 27: ...80 BPM with periodic right focus premature ventricular beats with 33 premature timing One Time event PVC 2 R on T Auto R on T Type 2 Premature Ventricular Contraction NSR of 80 BPM with periodic right...

Страница 28: ...six seconds followed by a repeating sequence of 2 atrial beats and 1 ventrical beat for six seconds Continuous Atrial Tach Atrial Tachycardia 160 BPM Continuous Paroxysmal Atrial Tach Paroxysmal Atri...

Страница 29: ...odic group of 11 Type 1 PVCs 11 PVC Type 1 36 NSR Continuous Run of 11 PVCs Man Run of 11 Premature Ventricular Contractions NSR of 80 BPM with periodic group of 11 Type 1 PVCs One Time event 6 PVCs p...

Страница 30: ...b Coarse Ventricular Fibrillation Irregular waveform with no real P wave or clear R R interval and a high signal level Continuous Vent Fib Fine Ventricular Fibrillation Irregular waveform with no real...

Страница 31: ...r four beats 160 220 400 470 ms followed by a P wave without a QRS Continuous 3rd Deg Heart Block Third Degree Heart Block 80 BPM with P wave rate of 80 BPM and QRS rate of 30 BPM Continuous Rt Bundle...

Страница 32: ...PS 2200 30 This page intentionally left blank...

Страница 33: ...Function Demand Occasional Demand Frequent AV Sequential Indicates Default Setting See Power Up Settings ECG RESPIRATION TEMPERATURE WAVEFORM ECG PACEMAKER None 50 Hz 60 Hz Muscle Wander Baseline Res...

Страница 34: ...Pulse at the start of each P wave Asynchronous Asynchronous Pacemaker Wave 75 BPM with Pacer Pulse at the start of each QRS wave and no P wave Non Capture Ventricular Pacemaker Wave with Periodic Non...

Страница 35: ...Wave 125 Hz Square Wave 2 Hz Triangle Wave 2 Hz Triangle Wave 2 5 Hz Sine Wave 0 1 Hz Sine Wave 0 5 Hz Sine Wave 5 Hz Sine Wave 10 Hz Sine Wave 40 Hz Sine Wave 50 Hz Sine Wave 60 Hz Sine Wave 100 Hz...

Страница 36: ...sired option Then is used to accept the new setting Alternately to see a submenu of all the options for a highlighted parameter use Use to scroll to the desired option Then is used to accept the new s...

Страница 37: ...performance waves starting with Square Wave 125 Hz incrementing at a fixed interval The interval may be set in the System Setup Menu under Auto Step Time A countdown timer is shown in the display The...

Страница 38: ...PS 2200 36 This page intentionally left blank...

Страница 39: ...t occurs There are 16 Blood Pressure settings available 17 static and 8 dynamic Each of the dynamic waveforms will synchronize with the NSR rate or arrhythmia selection Both an automatic and manual Sw...

Страница 40: ...tic 5 mmHg Static 0 mmHg Static 20 mmHg Static 40 mmHg Static 50 mmHg Static 60 mmHg Static 80 mmHg Static 100 mmHg Static 150 mmHg Static 160 mmHg Static 200 mmHg Static 240 mmHg Static 250 mmHg Stat...

Страница 41: ...e is selected the channel will automatically sequence through all of the Static Pressure settings starting with 10 mmHg incrementing at a fixed interval The interval may be set in the System Setup Men...

Страница 42: ...Step Time Auto Static All If Auto Static All is selected all channels will automatically sequence through all of the static pressure settings starting with 10 mmHg incrementing at a fixed interval Th...

Страница 43: ...mmHg BLOOD PRESSURE RESPIRATION ARTIFACT PS 2220 Arterial 120 80 Radial Artery 120 80 Left Vent 120 0 Right Vent 25 0 Pulm Artery 25 10 CVP 15 10 PAW 10 2 Left Atrium 14 4 Static 10 mmHg Static 5 mmH...

Страница 44: ...s selected the channel will automatically sequence through all of the Static Pressure settings starting with 10 mmHg incrementing at a fixed interval The interval may be set in the System Setup Menu u...

Страница 45: ...16 mmHg BLOOD PRESSURE RESPIRATION ARTIFACT PS 2210 Indicates Default Setting See Power Up Settings Available in NSR Only Arterial 120 80 Radial Artery 120 80 Left Vent 120 0 Right Vent 25 0 Pulm Art...

Страница 46: ...uto Static Pressure is selected the channel will automatically sequence through all of the Static Pressure settings starting with 10 mmHg incrementing at a fixed interval The interval may be set in th...

Страница 47: ...tep at a fixed time interval The sequence may be run on any channel while in the Normal Sinus Rhythm screen but only on one channel at a time The sequence can be activated by using to highlight the ap...

Страница 48: ...10 2 Deflate BLOOD PRESSURE 0 mmHg 1 mmHg 2 mmHg 3 mmHg 4 mmHg 5 mmHg 6 mmHg 7 mmHg 8 mmHg 9 mmHg 10 mmHg 11 mmHg 12 mmHg 13 mmHg 14 mmHg 15 mmHg 16 mmHg RESPIRATION ARTIFACT OPTION WAVEFORM SEQUENCE...

Страница 49: ...etup screen BLOOD PRESSURE 0 mmHg 1 mmHg 2 mmHg 3 mmHg 4 mmHg 5 mmHg 6 mmHg 7 mmHg 8 mmHg 9 mmHg 10 mmHg 11 mmHg 12 mmHg 13 mmHg 14 mmHg 15 mmHg 16 mmHg RESPIRATION ARTIFACT Set time per step 1 30 sec...

Страница 50: ...Mode during a sequence use The following message box will be displayed Right Atrium CVP 15 10 Right Ventricular 25 0 Pulmonary Artery 25 10 Pulmonary Arterial Wedge 10 2 Right Atrium CVP 15 10 Right V...

Страница 51: ...simulation can begin See SETUP for selection information There are 12 rate settings available 9 BrPM rates Apnea 0 BrPM and 3 timed Apneas The display will resemble the following 12 sec Apnea 22 sec A...

Страница 52: ...to scroll to the desired option Then is used to accept the new setting Timed Apnea To trigger a one time event Apnea for one of the specific periods 12 22 or 32 seconds simply highlight the desired p...

Страница 53: ...le the following These temperatures can be selected by using to highlight the parameter to change and using to scroll to the desired option Then is used to accept the new setting Alternately to see a...

Страница 54: ...PS 2200 52 This page intentionally left blank...

Страница 55: ...cing as well as whether the arrhythmias will display in order or randomly Subsets of the Arrhythmias can be selected to allow for individualization The PS 2200 is placed into TRAINING mode by pressing...

Страница 56: ...utomatically at the set time If in the manual mode use to go to the next arrhythmia when ready To pause or exit the Training Mode during a session use The following message box will be displayed None...

Страница 57: ...by marking those specific arrhythmias or groups of arrhythmias of interest After a subset of arrhythmias has been selected it will remain in memory It may them be edited at any time prior to starting...

Страница 58: ...display the list of arrhythmias for a specific group Then use to scroll through the arrhythmias Any arrhythmia marked with a check will be included in the subset Use to toggle the selection of an arr...

Страница 59: ...ias from that group use to return to the Group Submenu SOME will appear to indicate a partial selection of the arrhythmias in that group Additional groups may be modified in the same manor When done w...

Страница 60: ...PS 2200 58 This page intentionally left blank...

Страница 61: ...ion Baseline Ohms Respiration Lead and the Blood Pressure Sensitivity parameters These should be set according to the device under test Output Setup The Output Setup screen allows for the setting of t...

Страница 62: ...tware information The display will resemble the following These settings can be selected by using to highlight the parameter to change and using to scroll to the desired option Then is used to accept...

Страница 63: ...intensity of the backlight Note Lower intensities extend battery life 0 100 Battery Life Displays current life of the batteries At 5 a warning screen will appear At 10 the unit will power down automa...

Страница 64: ...PS 2200 62 This page intentionally left blank...

Страница 65: ...Output Disabled ECG Arrhythmia 1 0 mV Artifact None 0 0 mV ST Elevation Premature Atrial PAC Auto ECG Pacemaker Pulse Amplitude 5 mV Pulse Width 1 0 ms Atrial Waveform 1 0 mV 0 0 mV ST Elevation Artif...

Страница 66: ...and bring them back when the power is turned on Custom If this option is selected the user may save a unique set of default parameters and the unit will recall them every time the power is turned on...

Страница 67: ...the NSR BPM rate The output is off in Arrhythmia and Performance Modes The module plugs directly into the AUX 7 pin mini din connector and is powered from the PS2200 The output is only functional whe...

Страница 68: ...PS 2200 66 This page intentionally left blank...

Страница 69: ...veform to simulate uterine contractions The waveform may be triggered manually or set to occur periodically The Intrauterine Pressure IUP curve causes a change in the Fetal Heart Rate FHR based on the...

Страница 70: ...only selectable when not using the IUP Wave Simulation When IUP Simulation is active FHR is fixed at a base rate of 140 BPM and then varies based on the IUP wave Uniform Deceleration Early Deceleratio...

Страница 71: ...n The FHR goes from 140 BPM down to 100 BPM and back again The rate follows the 90 seconds bell curve of the contraction but is delayed by 45 seconds Uniform Acceleration The FHR goes from 140 BPM up...

Страница 72: ...PS 2200 70 This page intentionally left blank...

Страница 73: ...Monitor takes as a blood temperature input The PS 2200 drives an external Cardiac Output module This module MCO 2100 provides both a common connection point for the cabling and a manual simulation set...

Страница 74: ...ture 0 C or 24 C 2 Connect Com2100 Module to the CO Temp input on the PS 2200 3 Connect the Blood Temperature Sensor Line BT Thermistor Cable from the DUT to the small 4 pin connector on the COM2100 4...

Страница 75: ...to terminate the cycle Normal This test allows the selection of a flow rate 2 5 5 0 or 10 0 l min When run it will produce the normal temperature curve for the selected injectate temperature 0 C or 24...

Страница 76: ...PS 2200 74 This page intentionally left blank...

Страница 77: ...75 The following are representations of the socket connectors found on the unit They are viewed as if looking at the socket in the unit not the cable pins OUTPUT CONNECTIONS...

Страница 78: ...PS 2200 76 This page intentionally left blank...

Страница 79: ...HE SERVICE FOR WHICH THEY ARE INTENDED THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM THE DATE OF SHIPMENT EXCLUSIONS THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR IMPLIED INCLUDING B...

Страница 80: ...PS 2200 78 This page intentionally left blank...

Страница 81: ...ECG Performance Sine Wave 0 1 0 5 5 10 40 50 60 100 Hz Square Wave 0 125 2 000 Hz Triangle Wave 2 000 2 500 Hz Pulse Wave 30 60 120 BPM 60 ms Amplitude 05 10 15 20 25 30 35 40 45 50 1 0 1 5 2 0 2 5 3...

Страница 82: ...40 50 60 80 100 150 160 200 240 250 300 320 400 mmHg Accuracy 2 of Reading 2 mmHg Impedance 300 ohms Accuracy 10 Excitation Range 2 to 16 V RMS Excitation Frequency DC to 5 KHz Sensitivity 5 or 40 uV...

Страница 83: ...iform Acceleration IUP Wave Bell Curve with 90 mmHg Peak and 90 Sec Width IUP Trigger Manual Auto 2 3 5 Minute Cardiac Output Injectate Volume 10 cc Injectate Temperature 0 C or 24 C 2 Calibration Coe...

Страница 84: ...Inches 216 x 140 x 41 mm ABS Plastic WEIGHT 2 Lbs 0 91 Kg FACE PLATE Lexan Back printed OPERATING RANGE 15 to 40 C STORAGE RANGE 20 to 65 C Electrical POWER Battery 9 VDC 2 required NEDA 1604 Alkalin...

Страница 85: ...83 NOTES...

Отзывы: