22
The PS-2105 can send performance waveforms to ECG machines in 3, 5 or 12-lead
configurations. It has independent outputs for each signal lead, referenced to the right
leg.
There are 15 Performance waves available for testing and verifying. The PS-2105 is
placed into PERFORMANCE mode by pressing the category key.
The display will resemble the following:
These waves and amplitudes can be selected by using to highlight the
parameter to change and using to scroll to the desired option. Then
is used to accept the new setting.
Square Wave .125 Hz
Square Wave 2 Hz*
Triangle Wave 2 Hz
Triangle Wave 2.5 Hz
Sine Wave 0.1 Hz
Sine Wave 0.5 Hz
Sine Wave 5 Hz
Sine Wave 10 Hz
Sine Wave 40 Hz
Sine Wave 50 Hz
Sine Wave 60 Hz
Sine Wave 100 Hz
Pulse 30 BPM
Pulse 60 BPM
Pulse 120 BPM
Auto Wave
0.5 mV
1.0 mV*
1.5 mV
2.0 mV
* Indicates Default Setting
(See Power Up Settings)
AMPLITUDE
WAVEFORM
ECG
– PERFORMANCE