ARRHYTHMIA SEQUENCE: (NIBP-1030 Only)
Use the key to cycle through the Test Modes until Arrhythmia Sequence appears.
The purpose of this function is to provide a timed sequence of some of the more common
Arrhythmias. The test will continually cycle through the 6 arrhythmias in the sequence listed
below. A count down timer is displayed to indicate the time still remaining in each step.
NOTE: While in this mode, the NIBP simulation output is disabled.
The displays will resemble the following examples: (Typical Screens)
(Pressure Only Screen)
(Output Waveform Screen)
(Pressure Graph Screen)
Step
Display
Total Time
ECG Output
1
Single PVCs
18 sec
11 NSR, 1 PVC, 11 NSR, 1 PVC
2
Pair of PVCs
9 sec
10 NSR, Pair of PVCs
3
Single PVCs
18 sec
11 NSR, 1 PVC, 11 NSR, 1 PVC
4
Run of 5 PVCs
12 sec
11 NSR, Run of 5 PVCs
5
Vtach
30 sec
15 NSR, VTach
6
Vfib
30 sec
15 NSR, VFib
7
Asystole
30 sec
15 NSR, Asystole
PEAK PRESSURE
(Information not relevant to
this test, NIBP is disabled.)
Single PVCs
Pair of PVCs
Single PVCs
Run of 5 PVCs
Vtach
Vfib
Asystole
ARRHYTHMIA
PRESSURE
(Information not relevant to this
test, NIBP is disabled.)
PRESSURE ENVELOPE
OFFSET
This line is only displayed
if a non-zero value is
programmed into the
Systolic &/or Diastolic
Shift. (See Setup for
more information).
TIMER
Содержание NIBP-1010
Страница 1: ...NON INVASIVE BLOOD PRESSURE SIMULATORS NIBP 1030 NIBP 1020 NIBP 1010 USER MANUAL...
Страница 2: ......
Страница 31: ...23 Adult 120 80 with Pace Typical Screens Adult High 120 80 Typical Screens...
Страница 32: ...24 Adult Low 80 40 Typical Screens Neonatal 70 40 Typical Screens...
Страница 48: ...40 NOTES...