V.     APPLICATIONS 

 

The EndoCheck™ Test System is a versatile product. Several suggested applications and methods 
for using it follow. 

 

Suspected bad fiber cables 

Connect a known-good or new fiber cable of the same type as the suspect one, and take a reading 
with the light meter. Keeping the light source at the same setting, connect the suspect fiber cable and 
compare its reading to the good one. 
 

Suspected bad light source or lamp 

Connect a known-good or new fiber cable to a known-good light source, and take a reading with the 
light meter. Keeping the same fiber cable, replace with good light source with the suspect source and, 
at the same intensity setting, compare its reading to the good one. 
 

Fiber Cable Inspections as part of routine Preventive Maintenance (PM) Procedures 

Use EndoProjector™ to perform a visual inspection of fiber cables during each PM inspection. 
Record the technician’s assessment of relative amount of fibers that provide light throughput. 
 

Benchmarking Fiber Cables 

For each type or brand of fiber cables used in a facility, document the light meter’s reading for a new 
cable. For best consistency, always use the same light source at the same setting for all subsequent 
tests. Set a lower limit of light throughput based on this initial new-cable benchmark. For instance, if a 
new “Brand X” fiber cable reads 10,000 lux when connected to a specific light source at a 
predetermined intensity setting, you may choose to replace those fiber cables whenever they degrade 
to 80% of new light throughput, or 8,000 lux. (Note: These lux readings and percentage amounts are 
for illustration only, and are not recommendations. Each facility is encouraged to establish their own 
benchmarks based on their own unique situations and equipment.) 
 

Benchmarking Light Sources 

For each type or brand of light source or replacement light source lamp used in a facility, document 
the light meter’s reading for a new one. For best consistency, always use the same fiber cable and 
the same light source intensity setting for all subsequent tests. Set a lower limit of light output based 
on this initial new-source or new-lamp benchmark. For instance, if a new “Brand X” light source lamp 
reads 10,000 lux at a predetermined intensity setting, you may choose to replace those lamps 
whenever they degrade to 80% of new light output, or 8,000 lux. (Note: These lux readings and 
percentage amounts are for illustration only, and are not recommendations. Each facility is 
encouraged to establish their own benchmarks based on their own unique situations and equipment.) 
 

Other Uses for the Light Meter 

The light meter may be used for a variety of applications in the healthcare setting, including: 

 

X-Ray collimator light testing 

 

Surgery light testing 

 

General workplace lighting analysis 

 

Measurement of visible light for any purpose 

 

Do not exceed the sensor’s maximum capacity. See the warning elsewhere in this manual. 
 

 

Содержание ENDOCHECK Test system

Страница 1: ...User Manual ENDOCHECK Test System...

Страница 2: ...CONTENTS I INTRODUCTION 2 II SPECIFICATIONS 3 III RECEIPT INSPECTION RETURN 4 IV USE OF THE ENDOCHECK TEST SYSTEM 5 V APPLICATIONS 6 VI WARRANTY 7 VII TECHNICAL SUPPORT 7...

Страница 3: ...cal component that yields an approximately 10X magnification of the light signal coming out of the fiber cable The image when displayed on a wall or other flat surface generally allows the individual...

Страница 4: ...cal EndoProjector Construction Type Stainless steel body and precision glass optics Projection Magnification Approximately 10X Size 0 75 diameter x 1 375 long Light cable compatibility Adapters includ...

Страница 5: ...ument and provide the model and serial numbers If possible use the original packaging material for return shipping Otherwise repack the unit using a reinforced cardboard box strong enough to carry the...

Страница 6: ...t to allow counting of individual fibers or to make a quantitative assessment of cable quality It allows a quick visual and intuitive analysis of relative cable quality 7 If a quantitative analysis of...

Страница 7: ...ected to a specific light source at a predetermined intensity setting you may choose to replace those fiber cables whenever they degrade to 80 of new light throughput or 8 000 lux Note These lux readi...

Страница 8: ...ther form of guarantee concerning the EndoCheck Test System other than this guarantee Some jurisdictions do not allow disclaimers of expressed or implied warranties in certain transactions therefore t...

Отзывы: