3 Disassembly / Assembly
3 - 4
Infusomat® compact
plus
P 1.0
EN
For internal use, only
Zero force insertion connector
3.1
HOUSING FOOT
Designation Ord.
No.
Housing foot
Disassembly
1. Pull the four housing feet (
) out of the lower
part of the housing.
Assembly
1. Push the housing feet into the guides in the lower part of the
housing.
NOTICE
Make sure that the ribbon cable is centered between the guides
of the connector when the zero force insertion connector is
locked. Check the lock and that the ribbon cable is correctly
seated before continuing with installation.
Fig. 3 - 2
1 Housing foot
Содержание 8717070
Страница 1: ...Infusomat compactplus P Service Manual EN Version 1 0 For internal use only For internal use only...
Страница 14: ...Important Preliminary Remarks 0 14 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 16: ...Contact Information 0 16 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 24: ...1 Overview of the Device 1 8 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 32: ...2 Unit Diagnosis Adjustments 2 8 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 80: ...5 Servicing the Unit 5 2 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 84: ...6 Technical Safety Check TSC 6 4 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 94: ...7 Procedural Instructions on the TSC 7 10 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 100: ...9 Spare Parts List 9 4 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 105: ......